5 – Installation and set-up
MultiExact Gas Analyser
5.2.2 Analogue output connections
WARNING
The analogue output terminals are separated from the analyser
mains circuits by reinforced insulation. The terminals must only be
connected to circuits that are themselves separated from mains
voltages by an isolation method that provides at least this level of
protection.
CAUTION
To comply with EMC requirements, screened cables must be used to
connect the analogue outputs.
1.
Connect the wires in your cable to the screw terminals on the 6 pin connector
provided. Section 2.3 provides information on the rating and size of cable.
2.
The screen should be connected to the earth point on the rear of the analyser.
3.
Unless otherwise specified a mA analogue output is provided as standard
across pins 1 & 2.
Pin
Use
1
I+
2
I–
3
GND
4
V+
5
V–
6
GND
Figure 12 – Analogue output interface connector
4.
If your analyser is configured to provide voltage outputs, connect the wires in
your cable to pins 4 – 5 on the terminal strip.