SERVOPRO FID Online Total Hydrocarbons Analyser
SERVOPRO FID 01000001A rev:5 29
Here is the list of error messages that can appear in the historic menu. This table
describes every error message. For more information about an error, refer to
section 9.
Message
Meaning
Alarm #1
There is an alarm #1. If OK is written next to this message, the
error is resolved.
Alarm #2
There is an alarm #2. If OK is written next to this message, the
error is resolved.
Air Dev
Air flow deviation. If OK is written next to this message, the error
is resolved.
Air low
Low air flow. If OK is written next to this message, the error is
resolved.
Fuel Dev
Fuel flow deviation. If OK is written next to this message, the error
is resolved.
Fuel Low
Low fuel flow. If OK is written next to this message, the error is
resolved.
Smp Dev
Sample flow deviation. If OK is written next to this message, the
error is resolved.
FID T Dev
FID oven temperature deviation. If OK is written next to this
message, the error is resolved.
Flame Off
The FID flame is off. If OK is written next to this message, the
error is resolved.
Glow Plug
The glow plug is damaged. If OK is written next to this message,
the error is resolved.
IO Prob.
IO board communication problem. If OK is written next to this
message, the error is resolved.
Smp Prob.
Sample pressure regulator communication error. If OK is written
next to this message, the error is resolved.
Air Prob.
Air pressure regulator communication error. If OK is written next to
this message, the error is resolved.
Fuel Prob.
Fuel pressure regulator communication error. If OK is written next
to this message, the error is resolved.
RTD Prob.
RTD sensor is damaged. If OK is written next to this message, the
error is resolved.
Therm Pro.
Thermistor sensor for flame detection is damaged. If OK is written
next to this message, the error is resolved.
Table 19 : Error message and meaning
You can toggle between log pages pressing the
arrows. Refer to system
error section for error messages.
It is normal to see alarm messages when starting the analyser since
parameters like flow and temperature are not yet stable.