54 DF-760E
User Interface
7.5.2.3 AutoZero Setup
Use the
Next
key to move from between fields and use the arrow keys (
and
to
change the highlighted selections and to enter numerical values. When done, use the
Next
key to move to the
Accept
button and hit the
Enter
key to return to the main display.
Using the
ESC
key at anytime will exit the screen making no changes and return to the
main display.
Figure 27: Moisture AutoZero Setup Screen
Moisture AutoZero Setup
allows for unattended, automatic zeroing of the moisture
analyzer at preset times, preceded automatically by a purge of zero gas if desired.
If, when this screen is entered, and the
Scheduled AutoZero
is turned off, the box in the
upper right corner will not be displayed. When the
Scheduled AutoZero
is switched from
off to on, the box will appear and the time of the
Next
scheduled
AutoZero
will be set to
the current time and date. If no other changes are made, when this screen is accepted the
autozero process will start immediately. Use the
Esc
key to escape from the screen without
starting an autozero process.
Setting a time and date for the
Next AutoZero
is accomplished by accepting the current
time and date or using the
Next
key to move to the
Next AutoZero
box. With the left and
right arrow keys, move the cursor to the right of the digit you want to change and use the
up and down arrow keys set it to the desired value. When done, use the
Enter
key to move
to the next box.
NOTE: If the
Scheduled AutoZero
is then turned off, the autozero process is stopped and
returning to this screen will not show the
Next AutoZero
box in the upper right hand
corner.
The
AutoZero Interval
can be used to set the frequency of zeroing the moisture cell
starting with the setting in the box labeled Next AutoZero. The date and time of the next
scheduled AutoZero is automatically calculated based on the current AutoZero Interval
setting.
With the
AutoZero Timeout
the user can set the allowed time for stabilization (after
completion of any pre-zero purge). If the required stable period is not achieved during this
Содержание NanoTrace DF-760E
Страница 1: ...NanoTrace Dual Moisture Oxygen Analyzer DF 760E Instruction Manual...
Страница 4: ......
Страница 38: ......
Страница 44: ......
Страница 122: ......
Страница 128: ......
Страница 138: ......
Страница 142: ......
Страница 143: ...Service DF 760E 139...
Страница 144: ...140 DF 760E...
Страница 148: ......
Страница 166: ......
Страница 168: ......
Страница 171: ...DF 760E 167...
Страница 173: ...Appendix A User Menu Screens DF 760E 169 Page 65 Page 66 Page 67 14 2 Oxygen Menus Page 72 Page 74 Page 86 Page 87...
Страница 174: ...170 DF 760E Appendix A User Menu Screens Page 88 Page 91 Page 93 Page 97 Page 98 Page 99...
Страница 175: ...Appendix A User Menu Screens DF 760E 171 14 3 System Menus Page 102 Page 103 Page 103 Page 106 Page 108 Page 109...
Страница 176: ...172 DF 760E Appendix A User Menu Screens Page 110 Page 111 Page 113 Page 114 Page 115 Page 116...