Connecting to External Devices DF-760E
43
The maximum load resistance for each is 1K Ohms and the analyzer provides a compliance
voltage of approximately 28 VDC for each.
J4 Pin #
Moisture Signal
Description
J4-8
GND
Ground
J4-7
Key
J4-6
A Out +
Analog Voltage Output (+)
J4-5
A Out -
Analog Voltage Output (-)
J4-4
Unused
J4-3
Unused
J4-2
4-20 mA +
4-20 mA Output (+)
J4-1
4-20 mA -
4-20 mA Output (-)
Table 4: Pin-Out of Moisture Signal Output Connector J4
J4 Pin #
Oxygen Signal
Description
J10-8
GND
Ground
J10-7
A OUT +
Analog Voltage Output (+)
J10-6
A OUT -
Analog Voltage Output (-)
J10-5
Key
J10-4
Unused
J10-3
Unused
J10-2
4-20 mA +
4-20 mA Output (+)
J10-1
4-20 mA -
4-20 mA Output (-)
Table 5: Pin-Out of Oxygen Signal Output Connector J10
6.4
Relay Ports – J2, J3, J8, J9
Eight form C (SPDT) relays (contact closures) are provided to assign to the various alarms.
The contacts are rated at 30 VDC, 1A resistive load. They are not designed to switch AC
power.
The relay contacts can be programmed through the user interface for up to four oxygen and
four moisture levels, temperature, electrolyte level/condition, moisture cell diagnostics,
loss of gas flow, zero calibration in progress, moisture analyzer off line, freeze of analog
outputs during calibration.
The relays are wired for Fail Safe operation such that a Normally Open (No alarm) contact
connects to common when an alarm occurs or when power to the instrument is lost.
Relays 1, 2, 3, and 4 are assignable to moisture signals and relays 5, 6, 7, and 8 are
assignable to oxygen signals.
The relay wiring can be tested with the Relay test routine found on page 110.
Содержание NanoTrace DF-760E
Страница 1: ...NanoTrace Dual Moisture Oxygen Analyzer DF 760E Instruction Manual...
Страница 4: ......
Страница 38: ......
Страница 44: ......
Страница 122: ......
Страница 128: ......
Страница 138: ......
Страница 142: ......
Страница 143: ...Service DF 760E 139...
Страница 144: ...140 DF 760E...
Страница 148: ......
Страница 166: ......
Страница 168: ......
Страница 171: ...DF 760E 167...
Страница 173: ...Appendix A User Menu Screens DF 760E 169 Page 65 Page 66 Page 67 14 2 Oxygen Menus Page 72 Page 74 Page 86 Page 87...
Страница 174: ...170 DF 760E Appendix A User Menu Screens Page 88 Page 91 Page 93 Page 97 Page 98 Page 99...
Страница 175: ...Appendix A User Menu Screens DF 760E 171 14 3 System Menus Page 102 Page 103 Page 103 Page 106 Page 108 Page 109...
Страница 176: ...172 DF 760E Appendix A User Menu Screens Page 110 Page 111 Page 113 Page 114 Page 115 Page 116...