Appendix B – Hydrogen Service Safety System DF-760E
173
15 Appendix B
– Hydrogen Service
Safety System
The
Hydrogen Service Safety System
is designed to safeguard the DF-760 from
explosion hazards when operating on hydrogen sample gas under normal pressure and flow
conditions as described in this manual. The instrument chassis and the remote pump, if
equipped, are both protected by maintaining a safe condition within their respective
enclosures.
If equipped with this option at the time of shipment from the factory, the analyzer will be
set for Hydrogen operation through the GSF screen and the Hydrogen Safety Service
System will automatically be enabled.
15.1
Instrument
Air circulation fans
pull air in through the front door (and vent through the rear)
of the chassis. The fans are rated at 50 cfm (125 cfm max) which will maintain the
internal chassis space below the lower explosive limit (LEL)
1
.
Sample delivery interlock valve
blocks flow of sample gas from entering the
instrument chassis under various conditions as described below. It consists of a
normally closed pneumatically actuated UHP springless diaphragm valve
positioned on the sample inlet bulkhead.
Analyzer case purge
valve
introduces nitrogen purge flow into the chassis upon
closure of the sample delivery interlock valve. It consists of a normally open
pneumatically actuated valve which feeds a customer supplied and regulated source
of nitrogen purge gas through a bulkhead on the rear of the cabinet.
2
Instrument controls and logic
3
default, manage and actuate the above mentioned
components to maintain safe operating conditions.
15.2
Vacuum Pump
If equipped with a vacuum pump,
an enclosure
equipped with a nitrogen case
purge is provided to maintain a safe condition. In addition, an interlock is provided
to detect and react to loss of Nitrogen purge flow. See Figure 146 on page 178.
15.3
Installation
If equipped with this safety system, the analyzer installation procedure is modified as
follows:
If the analyzer has the optional vacuum pump, the pump enclosure must be
mounted to a nearby wall or inside the rack shared by the analyzer. See Figure 146
on page 178.
Содержание NanoTrace DF-760E
Страница 1: ...NanoTrace Dual Moisture Oxygen Analyzer DF 760E Instruction Manual...
Страница 4: ......
Страница 38: ......
Страница 44: ......
Страница 122: ......
Страница 128: ......
Страница 138: ......
Страница 142: ......
Страница 143: ...Service DF 760E 139...
Страница 144: ...140 DF 760E...
Страница 148: ......
Страница 166: ......
Страница 168: ......
Страница 171: ...DF 760E 167...
Страница 173: ...Appendix A User Menu Screens DF 760E 169 Page 65 Page 66 Page 67 14 2 Oxygen Menus Page 72 Page 74 Page 86 Page 87...
Страница 174: ...170 DF 760E Appendix A User Menu Screens Page 88 Page 91 Page 93 Page 97 Page 98 Page 99...
Страница 175: ...Appendix A User Menu Screens DF 760E 171 14 3 System Menus Page 102 Page 103 Page 103 Page 106 Page 108 Page 109...
Страница 176: ...172 DF 760E Appendix A User Menu Screens Page 110 Page 111 Page 113 Page 114 Page 115 Page 116...