128
© Servomex Group Limited. 2020
07931001B revision 6
The flange bolt pattern must be:
4 bolt pattern
45 °
8 bolt pattern
22.5 ° (not shown)
Figure 7-4: Process flange positioning tolerance
Note:
Failure to align the transmitter and receiver flanges may adversely affect the
performance. Tolerance to misalignment is dependent on divergence, path length
and measurement. For further advice contact Servomex or local representative.
Mounting rigidity
Make sure that the flue wall is strong enough to support the weight of the
Laser 3 Plus transmitter and receiver units, including the mounting
flanges and unsupported cables.
Install the Laser 3 Plus on a structure that is capable of supporting the weight of the
transmitter unit (TU) and receiver unit (RU). If the structure is not strong enough,
you will experience alignment problems during commissioning and operation.
The following illustrations of a high temperature mounting arrangement for the TU
and RU, where the units have been mounted 200 mm (
7.87”
) from the process wall.
Particular attention is required on the TU as the extra weight may require
reinforcement of the process wall to counter the moment produced by the distance
from the process wall.
Always allow for the weight of the cable in addition to the weight of the TU and RU.
Acceptable
Not acceptable
Содержание 07931001B
Страница 1: ...Part Number 07931001B Revision 6 Language ENGLISH UK US SERVOTOUGH Laser 3 Plus OPERATOR MANUAL...
Страница 2: ...Servomex Group Limited 2020 This page intentionally blank...
Страница 4: ...Servomex Group Limited 2020 This page intentionally blank...
Страница 12: ...viii Servomex Group Limited 2020 07931001B revision 6 This page intentionally blank...
Страница 14: ...x Servomex Group Limited 2020 07931001B revision 6 This page intentionally blank...
Страница 172: ...158 Servomex Group Limited 2020 07931001B revision 6 Figure 7 25 Installation ball joint adjustment screws...