Quick Start Guide
SERVO
PRO CHROMA/NANOCHROME/ULTRA Gas Analyser
16
04400003A_0
When the hard drive is 99% full, saved data will be over-written with
the new data. The oldest data will be over-written first.
Note that a tick mark (little vertical line) is displayed on the real-time
chromatogram at every starting and ending peak to help identify the integration
windows.
V
iew each conditioning board’s trending indep
endently by clicking on the
corresponding tab at the top of the chromatogram.
1.6.10 Undo Zoom (F8)
Use the mouse to zoom into the trending. Press and hold the left mouse button
while moving the mouse. A rectangle will appear on the screen to identify the
zone to zoom. Releasing the left mouse button will apply the zoom. Press the
Undo Zoom
button to show the original trending without any zoom.
1.6.11 Print (Alt F2)
The
button enables printing of the chromatogram. To add a printer, see the
CONFIGURATION-PRINTERS
menu
.
1.6.12 Remote Countdown
A remote countdown field shows the time left when remote starting is activated.
The remote time can be set in the
CONFIGURATION-SYTEM
menu
Note: If the analyser has multiple configurations, the name of the current
configuration will appear in the bottom right corner of this menu.
1.7 Historic Menu
1.7.1 Alarm Menu
Displays the last 200 system alarms, peak value alarms and event logs, and all
the active alarms.
1)
System alarms:
•
Low sample flow
When the sample flow goes below 10 cc/min with a sample flow set point
greater than 10 cc/min.