Servomex 2500 Quick Start Manual
30
2500 Series - Configuration and Operation 02500/003E/0
5
VIEWING CONFIGURATION AND HISTORY FILES
5.1 Displaying ‘current’ alarms and faults
• Alarms
Go via MENU/ALARMS to DISPLAY ALARMS - Press ENTER
At display screen
Operation
NO ALARMS DETECTED
OR
[Measurement] AL1
100.0 vpm
HIGH ↑
Either ‘NO ALARMS DETECTED’ OR the first alarm is
displayed; if further alarms are present, an arrow will be
shown; use the cursor keys to access them
• Faults
Faults on the 2500 series are categorised into 2 types:
i)
General Faults
These indicate that a parameter is out of tolerance and user intervention is
required as soon as possible to correct any hardware problems arising or deal
with any failure of auto calibration, sample flow, etc.
ii)
Serious Faults
If any fault arises, or parameter degrades to such an extent that the
measurement accuracy is also significantly degraded, the 2500 goes into
‘Shutdown state’. The measurement is automatically withdrawn and all analogue
outputs are driven HIGH (or LOW, as configured; see installation manual). The
display will show a ‘Measurement Invalid’ message instead of the measured
value.
For full details of types of fault and the recommended remedial actions, see the
installation manual.
Go via MENU/FAULTS to DISPLAY FAULTS - Press ENTER
At display screen
Operation
NO FAULTS DETECTED OR
DET
SIG LO ↑
Either ‘NO FAULTS DETECTED’ OR the first fault is
displayed; if further faults are present, an arrow will be
shown; use the cursor keys to access them
Содержание 02500003E
Страница 1: ...Part Number 02500003E Revision 0 Language UK English SERVOTOUGH SpectraExact 2500 QUICK START MANUAL...
Страница 2: ...Servomex 2500 Quick Start Manual 2 2500 Series Configuration and Operation 02500 003E 0...
Страница 3: ...Servomex 2500 Quick Start Manual 2500 Series Configuration and Operation 02500 003E 0 3...
Страница 4: ...Servomex 2500 Quick Start Manual 4 2500 Series Configuration and Operation 02500 003E 0...