
SOL-96.110.041 / 290621
TOPAZ Total Alkalinity
Maintenance
39
6.6.1
Manual Measurement
The <Manual measure> function can be used to verify the instrument
performance using a standard solution. It is recommended to per
-
form one or two manual measurements before each calibration to
obtain a meaningful result.
1
Start the manual measurement by pressing the button.
2
When the message "Add standard" appears, fill 2/3 of the
calibration vessel [A] with the standard.
3
Wait until the calibration vessel is empty and the message
“Add standard” appears again.
4
Repeat steps 2 and 3 until the measurement starts automatically.
5
When the measurement is finished, add standard solution to the
calibration vessel again as requested on the screen.
6
At the end of the manual measurement, a window displays
the result of the current and the previous (“Bf”) manual measure
-
ment. Confirm by pressing the button.
Note:
The result of the latest manual measurement can later be
viewed in the parameter menu.
A
Calibration vessel
A
MANUAL MEASURE RESULTS
Manual Measure(Bf )
Manual Measure
9.9
10.0