Figure 5: Result visualization of impedance measurements
ment results. In the
strict mode
, the measurement is classified as NOT PASSED
if the tolerance mask is violated at any point. In the
non-strict mode
, the
measurement is classified as NOT PASSED if the average impedance result is
out of the limits. The mode can be set in the Seunis.config file (see section
) by setting option
Pass-fail strict
either to
"1"
or
"0"
.
•
The tolerance mask can be turned on/off by using the
button in
the tool bar. This also effects displayed impedance results (average &
min./max. impedance) and the small LED indicators in the status bar.
The indicators are grayed out if the tolerance mask is switched off.
•
The trace shown in the
Result Visualization Window
can be set to im-
pedance or step response via the menu entry
View
→
TDR
.
•
The x-axis can be changed from time [ns] to a space [cm] visualization
via the
View
→
TDR
→
x-axis
menu entry in the
Result Visualization
Window
. Make sure a
Reference Measurement
is performed to move