3 Maintenance/Safety & Functionality Tests
SenTec Digital Monitoring System
41
3.3
Docking Station
3.3.1
Cleaning and Disinfection
Clean or disinfect the Docking Station (including the Docking Station gasket) as often
as required per institutional ordinances (recommended weekly) using a cotton swab
moistened with 70% isopropanol.
3.3.2
Visual Inspection
Inspect the Docking Station door and the gasket for mechanical damages. The closing
mechanism must lock properly.
If you detect a possible damage, replace the defective part according to
section
Note: We recommend changing the gasket of the Docking Station annually.
Figure 10 Correct orientation of the Docking Station Gasket
3.3.3
Gas Level Indicator
Tools
Full
Service Gas bottle
Check the functionality of the gas level indicator using a new gas bottle.
1.
Screw in a new gas bottle and switch the SDM ON.
2.
Insert the sensor in the Docking Station and verify that the "100% gas icon"
displays in the status bar.
3.
Remove the gas bottle from the SDM and verify that the yellow "0% gas icon"
displays in the status bar.
If the expected result is not observed, refer to
section
troubleshooting" to further evaluate the problem.
Содержание SDM
Страница 1: ...For the SenTec Digital Monitoring System Software Version SMB V08 02 MPB V06 02 ...
Страница 6: ......
Страница 28: ...3 Maintenance Safety Functionality Tests HB 005615 j 22 Figure 6 Examples PCO2 and PO2 Slope Test ...
Страница 84: ...5 Repairs of the SDM HB 005615 j 78 Figure 17 Positioning of the Docking Station Gasket ...
Страница 93: ...8 Appendix SenTec Digital Monitoring System 87 HB 005615 j Date of release 02 2019 ...