Alarm Setpoints:
Use this box to programme the set-point value at which
you want the alarm to activate. The units are in %, ppm or mbar depending
on which sensor you have assigned to the alarm.
Once you have finished editing the parameters, select ‘Write Configuration to
Analyser’ from the file menu and the new data set will be written to the Rapidox.
The Rapidox is reprogrammed using the new configuration. The software then
reads the new settings back from the Rapidox and displays them in the boxes on
the left-hand side confirming that the configuration was successful. When you
have finished programming the analyser for new calibration gases you must
recalibrate the instrument to benefit from the changes.
8.6
On-Screen LCD
It is possible to display an LCD emulator on the desktop of your PC. This
reproduces exactly the display on the Rapidox unit and is convenient if you are
some distance away from the machine. To activate the LCD select the UTILITIES
menu and select the LCD on option. This will display a small window showing
the LCD that can be positioned independently anywhere on the desktop.
Figure 13: On-screen LCD. Note that the second line of the LCD is switched off in Heller
versions.
The blue LED indicator will flash to show that the display is updating.
8.7
Remote Calibrating and Cleaning
Before calibrating please read section 7.6 carefully taking particular note on the
guidance about calibrating with pressurised gas from a cylinder.
With the on-screen LCD display active it is possible to remotely calibrate the
analyser and clean the sensor. Click on the down arrow icon in the bottom right-
hand corner of the LCD display to expand the window: