Cambridge Sensotec Rapidox 1100ZR-H-PFC-E O
2
Analyser Instruction Manual
Version 1.2
15
Last printed 22/07/2015 18:08:00
7.10
The notation used by the display is fully programmable to suit your needs using
the keypad (2) or the software provided (section 8.4). The following options are
available to you:
1.
Oxygen:
The oxygen reading can be displayed in percent (e.g. 20.95%),
ppm (e.g. 0.01ppm) or scientific (e.g. 2.095E+05ppm). If percent is selected
the Rapidox will display oxygen as a percentage down to 0.1% and ppm
below this value. There are also options to display the oxygen reading in
terms of a pressure (units will be the same as the units selected for
pressure) or a nitrogen balance (%N
2
) reading. In this case the reading
displayed is 100-O
2
% =N
2
so for example in air 21% the nitrogen balance
reading will display 79% N
2
. Note that this is NOT an actual nitrogen
reading.
2.
Pressure:
The reading can be displayed in mbar, bar, psi, torr or kPa.
To program these options scroll down to menu no. 5 “Set Units” and press ENT.
Use the ENT key to progress through the list and the UP & DOWN keys to select
the required units. Once selected press ENT to store and return to run mode.
7.11
Setting the Display Options
The user can select the LCD refresh rate and the format of the line 2 display using
the keypad or the software provided. Scroll down the menu to option no. 6 “Set
up Display” and press ENT. Use the UP & DOWN arrows to select either the LCD
refresh rate or Line 2 Options:
1.
LCD Refresh Interval:
The frequency of the display update can be selected
from 0.1 to 1.5 seconds. The current setting can be edited using the same
method as described previously. Once edited press ENT to store and return
to the main menu. The default setting is 0.3 seconds.
2.
Line 2 Options:
The information displayed on line 2 of the LCD can be
chosen by the user. The options are None or Pressure (displayed in mbar,
bar, psi, torr or KPa. Use the UP & DOWN arrows to select the required
option and press ENT to store and return to the main menu. Note that on
the Heller coded machines the second line display is switched off by
default.
7.12
Pressure Mode
The user has the option of using the reading from the internal pressure sensor to
automatically correct for changes in gas pressure and hence oxygen pressure. A
manual pressure correction value for ambient pressure conditions can also be
entered to improve accuracy of the correction (factory setting is 1013 mbar