Cambridge Sensotec Rapidox 1100ZR-H-OL O
2
Analyser Instruction Manual
Version 1.2
2
Last printed 22/07/2015 18:02:00
Optional printer attachment
3
Heller Specific Features
Internal carbon trap fitted to the sample gas line
Heller model name on the boot up screen
Voltage input on rear panel to enable / disable the sample pump
Modified software
4
Technical Specification
Property
Specification
Supply Voltage
90-260VAC, 50/60Hz
Power consumption
100W (max)
Analyser dimensions
250mm X 263mm X 150mm
Weight
3.5 kg
Display
16
×
2 character (9mm) back lit LCD
Warm up time
3-4 minutes at 20°C
Normal operating temperature
5-35°C
Sensor Range
1ppm to 100% O
2
Sensor Accuracy
±
1% of the actual measured oxygen content
Outputs:
O
2
& pressure
0-10V (user-programmable) into minimum
5k
Ω
O
2
& pressure
4-20mA current loop (user-programmable)
into maximum 500
Ω
O
2
high and low alarms
Relay circuits – fully user-programmable
All data and parameters
RS232 or RS485 - data streamed on demand
Sample Pump
Mains type vacuum pump (variable flow)
with power switch on rear panel & voltage
input enable / disable
Flow Rate
0-1.5 L.min
-1
Max. gas temperature on input
60°C
Calibration
Requires 2 or 3 user-selectable gas
compositions (air is default plus another two)
Table 1: Rapidox 1100ZR-H-OL technical specification
5
Precautions
Make sure you read and understand these instructions and keep them safe for later
use. The unit should not be exposed to extreme temperatures < -5°C or > 60°C.
Normal operating temperature is 5-35°C. Avoid direct sunlight. Do not use liquid
cleaners, aerosols or solvents to clean the case. Use a damp cloth for cleaning. Do
not use this equipment near water. Avoid touching the LCD display as this may