
Cambridge Sensotec Rapidox 1100ZR-H-OL O
2
Analyser Instruction Manual
Version 1.2
23
Last printed 22/07/2015 18:02:00
columns) on any of the ‘Config’ pages in order to transfer the data into the edit
boxes, select ‘Save Configuration as’ from the ‘File’ menu and choose a filename.
This data set can be subsequently reloaded at any time by selecting ‘Load
Configuration File’ from the ‘File’ menu and then selecting the folder where the
configuration files have been stored.
If at any time you wish to restore the machine to its factory default settings select
‘Default Configuration’ from the ‘File’ menu to load these values. The software
will search for a unique ‘.rxc’ file that is loaded onto the PC during the software
installation. The filename is 2100***.rxc where *** is the last three digits of the
serial number located on the rear of the machine. This file is also located on the
memory stick that came with the machine. Normally this file is located in “My
Documents” directory but if the software cannot locate it then it will ask you for
the serial number to help it search. With the correct rxc file loaded into the ‘Read’
columns, simply click ‘>>’ (copy) to transfer the defaults into the edit boxes and
then click any of the three ‘Write’ button to load them back into the Rapidox.
The right-hand set of white editable text fields (‘write fields’) is used to enter new
values to be programmed into the Rapidox. The values in these fields can either
be entered manually, or copied across from the ‘read fields’ by clicking the vertical
‘Copy’ button, and then edited as required, however greyed boxes under the
‘Write’ buttons are not editable as they contain calibration data; they can only be
changed by copying data across from the ‘Read’ fields. Note that when you click
on the copy button all three ‘Config’ pages are updated at the same time and there
is no need to repeat the action when you toggle to the other page. Alternatively,
you can load saved configuration data by selecting ‘Load Configuration’ from the
‘File’ menu and selecting the required file. In order to programme the analyser
select ‘Write configuration to the analyser’ from the ‘File’ menu, or click the ‘Write’
button on any ‘Config’ page.
8.5
Reconfiguring the Analyser
The on-screen edit boxes contain variables that can be reprogrammed into the
Rapidox in exactly the same manner as using the front panel keypad. For a full
description please refer to the relevant section in this manual.
The user can choose whether to use ppm or percent notation to edit the text boxes,
where appropriate, by selecting the ‘UNITS’ menu and clicking on ‘Percent’ or
‘ppm’. To access a field, either click inside it or use the TAB key to scroll through
them. The field parameters are as follows:
8.5.1
Config 1 Screen
1)
Oxygen Range:
Not available for the Rapidox 1100ZR-H-OL
2)
Oxygen Units & Format:
See Section 7.11 for a full description. Use the
drop down box to select PPM EXPONENTIAL, PPM MIXED, AUTO%/PPM,