Cambridge Sensotec Rapidox 1100ZR-H-OL O
2
Analyser Instruction Manual
Version 1.2
20
Last printed 22/07/2015 18:02:00
held down, the printer will keep printing data (but not the serial number).
When the printer is first connected or turned on, a “start-up” message like the one
displayed on the LCD when the Rapidox is switched on will be printed, and will
look like this:
Cambridge
Sensotec Ltd
Rapidox 1100ZR-H-OL
Oxygen Analyser
S/N: 2100942
FW v 01.03.03.01
7.17
Load Defaults
If you make a mistake programming the Rapidox it is possible to restore the
machine back to its factory settings by loading the unique default configuration
that is supplied on your software memory stick. The default set-up can only be
accessed using the software provided. Please see section 9.23 for further details.
Alternatively, you can reset to ‘generic’ factory defaults directly from the front
panel keypad. Press ENT to enter the user setup menu and select menu option 10
“Load Defaults” and press ENT to proceed. The next screen will say “Load
Defaults ?” with the prompt to press and hold ENT for 2 seconds to confirm. To
complete this press the ENT button until the bar graph has completed and the
screen says “Defaults Loaded” This will give you all the basic settings that are
factory set. You will now need to re-calibrate the sensor and check all the settings
are suitable.