Electrical Warning
A professional licensed electrician should perform all
electrical work/installation.
•
Installation Requirements
1.
SpecMetrix traversing systems must be
installed inside a cage.
2.
Indoor and dry location use only
3.
Environmental temperature: 0-50C (32-122F)
4.
System mounting elevation: Max 2m (6.5ft)
5.
The weight of total system ~1000lbs (453kg, ICU weight 125lbs)
System Specifications
Performance Specifications
1.
Coating Thickness Range
0.5 to 250 microns
2.
Typical Accuracy
3%
3.
Non-contact measurement range
15-33mm
Environmental Specifications-Customer must provide:
•
Dedicated Circuit
•
Single-Phase with neutral and ground
•
50/60Hz
•
120/240VAC
•
15Amp Circuit (min 14AWG (1.63mm
diameter wire)
•
SpecMetrix ICU can be hard-wired to the
main power input or a wall socket
•
Primary Operator Workstation will require
one or two wall sockets depending on the
configuration that must be in the same
dedicated circuit
•
All Ethernet cables must be minimum CAT 6
Shielded
•
ICU can be maximum of 65.5 feet (~20 meters) “cable-routing” distance away from actuators
•
Ethernet cables with a distance more than 45.72m/150ft
MUST
be converted to an armored
fiber optic multi-mode
•
Highly Recommended to use a 1000VA or better UPS.
1.
SpecMetrix® Display Unit
6 amps @ 120VAC, 4 Amp @ 240VAC
2.
Operating Humidity
10%-90% relative, non-condensing
3.
ICU enclosure protection
IP65/66
4.
SpecMetrix® display protection
IP65/66