METIS M311 / M322 / H311 / H322 (17-pin)
Electrical Connection
10
M3 / H3
0...20 mA
4...20 mA
A
0...20 mA
4...20 mA
A
+I
OUT1
-I
OUT1
+I
OUT2
-I
OUT2
Green (+)
Yellow (-)
Red (-)
Blue (+)
Note:
The outputs are galvanically
isolated from the supply voltage.
Max. 100 mA
0 V
Min. 240
Ω
M3 / H3
Output
+24 V
100 k
Ω
White
Grey-pink
Brown
0 V
Red-blue
4.1.2 Analog Output 1 and 2
The analog outputs are adjustable (e.g. for a temperature display
device) to:
0
–20 mA
or
4
–20 mA
Furthermore, the following adjustments can be done:
Analog output 1
always provides the measured temperature
(always the temperature selected and displayed in the control
window in
SensorTools
Control window
in
SensorTools
).
Analog output 2
can be assigned to provide different signals:
-
2-color temperature
-
1 channel temperature (optional channel 1 or 2)
-
Device temperature
-
Control output (= manipulated variable) for PID controller
4.1.3 Digital Inputs / Outputs and Analog Input
7 ports are each available as (configuration under
Device settings and configuration
):
Digital output of a switching signal (invertible, see under
Various settings
to:
-
Device ready to operate (device ready and error-
free after self-test)
-
Material detection (limit switch turns on when ex-
ceeding the beginning of temperature range)
-
Limit switch when a certain temperature threshold
is exceeded or falling below
-
Exceeding the maximum allowed device tem-
perature
-
Signal strength too low (as dirty window alarm
function: enables to detect the degree of contamination of the pyrometer‘s optics, viewing window
or identify interferences (dust...) in the IR sensor’s sight path and trigger an alarm if necessary).
-
Controller activity of PID controller, control process within adjustable setpoint limits, control pro-
cess successfully completed, hold time finished.
Digital input:
A voltage pulse of 24 V (exception: continuous voltage at "Activate controller"), pulse
length adjustable via software (minimum 3 ms, factory settings 100 ms) under
enables:
-
-
Start / stop controller (PID controller, see
).
-
Activate controller: The control process is activated
as long as voltage at the terminal is present.
-
Load (stored) pyrometer configuration (also PID
controller parameters) (only digital inputs 1-3, see
under
).
-
Switch on / off laser targeting light (when equipped
with laser targeting light).
(the targeting light will automatically switched off after 3 minutes if it is not switched off manually.
Adjustable via software under
Laser targeting light settings
).
CAUTION, Laser Radiation, Laser Class 2:
Never look into the direct or reflected laser beam.
Do not point the laser to anyone.
If laser radiation hits the eye, the eye should be averted immediately.
>10 kOhm
0 V
+24 V
Push
button
0 V
+24 V
White
Brown
Black
Violet
Pink
Grey