METIS M308 / M309 / M313 / M316 / M318 / M323 / H309 / H316 / H318 (12-pin)
SensorTools Software
49
7.2.6 Setups
Up to 7 different parameter settings (setups) can be stored in the pyrometer and activated via digital in-
puts (i.e. via external switch), (see
Configurable inputs / outputs
). Depending on which and how
many setups are to be used, one, two or all three inputs have to be used (see under
dow
The parameters for the selected setup can be entered directly in this window using the keyboard, with
Enter or the Tab key they are transferred to the pyrometer.
Setups can be saved individually or as a complete list to the PC and loaded from there.
Caution:
With an active Setup, the pyrometer operates using the setup
values. All the adjustment in the pyrometer software window are hid-
den.
Active setups are displayed:
At least 1 digital input is set to "Parameter selector", but no input
is activated (inputs to "low").
At least 1 digital input is set to "Parameter selector" and at least 1
input is active ("high").
Selecting of the setup
to be edited.
Enter parameters of
the selected setup
(changes are saved
directly into the py-
rometer with the "En-
ter" key or tab key.
Save current setup
e.g. on your hard
disk, or load saved
back up setup
Pyrometer parameters
Controller parameters
(only with PID control-
ler equipment)
Take over setup values from
the control windows in this
window, or replace parameters
in the control window with
those from this window
Save setup list (all 7
setups) or load setup
list to into the py-
rometer
Active setup display
M309-7682