METIS M308 / M309 / M313 / M316 / M318 / M323 / H309 / H316 / H318 (12-pin)
Configuring the Pyrometer
24
Status LEDs for devices with Profibus / Profinet
Operation mode LED
Meaning
Off
Not online / no power
Green
Online, data exchange
Flashing green
Online, clear
Flashing red (1 flash)
Parameterization error
Flashing red (2 flashes) Configuration error
Module status LED
Meaning
Off
Not initialized
Green
Initialized
Flashing green
Initialized, diagnostic
event(s) present
Red
Exception error
Module status LED
Meaning
Off
Not Initialized
Green
Normal operation
Green, 1 flash
Diagnostic event(s)
Red
Exception error
Red (combined with a
red network status LED)
Fatal event
Alternating red / green
Firmware update
Network status LED
Meaning
Off
Offline
Green
Online
(run)
Green, 1 flash
Online
(stop)
Green, blinking
Blink
Red (combined with a
red module status LED)
Fatal event
Red, 1 flash
Station name error
Red, 2 flashes
IP address error
Red, 3 flashes
Configuration error
Link / activity LEDs
Meaning
Off
No link
Green
Link
Green, flickering
Activity
6.2
Key Functions
Parameter:
Press the button repeated to get access to all settings sequentially (see
Pa-
rameters / Settings
). If a subcategory is found, this is indicated with a > (e.g. MEASPARA.>).
Pressing ENTer the subcategories are called.
Arrow keys:
Settings can be done with the arrow keys. Keep the button pressed to change
the values faster, reasonable for example when changing a temperature value. A changed
value is shown by a blinking display and will only be saved when pressing “Enter”, pressing
PAR calls the next parameter point.
If 60 seconds of inactivity have been made, the display jumps back to the measurement dis-
play without taking over any modified values.
Enter:
opens a parameter category or takes over a modified parameter value.
Key lock:
Press the arrow keys and the ENTer key simultaneously for approx.
3 s to activate or deactivate the key lock. With locked keys, no parameters can
be called and changed.
Laser targeting light
(depending on configuration): Turns on and off the laser targeting light.
CAUTION, Laser Radiation, Laser Class 2:
Never look into the direct or reflected laser beam.
Do not point the laser to anyone.
If laser radiation hits the eye, the eyes must be intentionally closed and the head
immediately moved out of the beam.
Profinet devices
Profibus devices
Not used
Operation mode
Modul status
Not used
Module status
Network status
Link / activity LED for A
Link / activity LED for B
A
B