System Configuration
111
Figure 6-6: Setting the Sampling Interval
Press the
+
and
–
buttons to toggle between
Normal
and
High
. Selecting High will double the
sample points if you need to have higher resolution of the reflected wave. Note that this also
increases data volume and may slow down data collection.
It is rare for the user to change this value, unless they have specific requirements.
The following table displays recommended temporal sampling interval for different frequencies:
Center Frequency
(MHz)
Recommended
Sampling Interval (ps)
12.5
6400
25
3200
50
1600
100
800
200
400
250
400
500
200
1000
100
6.1.5
Stacks
In noisy environments, one way to improve the signal to noise ratio is to collect more than one
trace at each position, average them, and then record the average trace; this is known as
Содержание pulseEKKO
Страница 1: ...2005 00040 09 ...
Страница 2: ...ii ...
Страница 4: ...Warranty Confirmation iv ...
Страница 10: ...x ...
Страница 14: ...Introduction 2 ...
Страница 72: ...Assembling Full Configurations 60 ...
Страница 80: ...Getting Started 68 ...
Страница 116: ...Tools and Preferences 104 ...
Страница 180: ...Collecting Data 168 ...
Страница 185: ...Slice View 173 Figure 8 4 Varying the Grid Lines from top to bottom ON PARTIAL and OFF settings ...
Страница 190: ...Map View 178 ...
Страница 198: ...Managing Data 186 ...
Страница 221: ...Care Maintenance 209 ...
Страница 222: ...Care Maintenance 210 ...
Страница 226: ...Technical Specifications 214 ...
Страница 228: ...Appendix A 216 ...
Страница 234: ...Appendix B 222 ...
Страница 236: ...Appendix C 224 ...
Страница 238: ...Appendix D 226 ...
Страница 246: ...Appendix E 234 ...
Страница 248: ...Appendix F 236 ...
Страница 250: ...Appendix G 238 ...