Страница 2: ...p Test 4 RUN HOLD Mode 4 Viewing the Measurement 5 Power Off 5 Automatic Backlight 6 Battery Replacement 7 Cleaning 7 High Voltage Operation 7 Troubleshooting 8 Warranty 10 Quality Assurance Back Cover Available Stock Codes 8 020 XT 50HZ 8 020 XT 60HZ 8 020 XT EURO 8 020 XT FRG 8 022 50HZ 8 022 60HZ 8 022 EURO 8 024 50HZ 8 024 60HZ 8 024 EURO ...

Страница 3: ... 0 500kV 0 69kV 0 400kV Current 1 5000A 1 2000A 1 2000A Sensor Opening Standard Wide Jaw Wide Jaw Opening Width 2 5 in 6 35 cm 3 86 in 9 8 cm 3 86 in 9 8 cm Type of Reading Single Single Single Resolution Amps 1 99 9A 0 1A 0 1A 0 1A Amps 100 1999A 1A 1A 1A Amps 2000 5000 0 01kA N A N A Weight 2 15 lbs 0 97 kg 3 8 lbs 1 76 kg 3 8 lbs 1 76 kg Accuracy 1 2 Counts Frequency Actual frequency indicated ...

Страница 4: ... considered the sole voltage isolation device For safety purposes the face plate battery cover chuck and entire Ampstik should be considered to be at the same potential Putting the face plate battery cover chuck or other parts of the Ampstik within the air gap of adjacent phases or ground could cause a phase to phase or phase to ground fault WARNING Do not force the arms of the Ampstik Plus ...

Страница 5: ...t changes The unit is immediately in the RUN mode after powering on To place the instrument into the RUN mode from a different mode press and hold the control switch and scroll to the option RUN then release the control switch to engage the option To take measurements with the True RMS Ampstik Plus in the RUN mode place the conductor between the two arms and observe the display For maximum accurac...

Страница 6: ...ARING THE DISPLAY Clearing the stored measurements The True RMS Ampstik is now returned to the RUN Mode Push and release the switch to switch between RUN and HOLD Modes POWER OFF Turning the instrument OFF Press and hold the control switch to scroll to the OFF option Release the control switch The True RMS Ampstik will turn itself off automatically after several minutes of inactivity Page 5 Operat...

Страница 7: ...IGHT Manually power off the backlight Press the control switch until the following screen appears After releasing the switch the display will momentarily show The backlight will stay off until the user manually powers it back on POWER ON BACKLIGHT Manually power on the backlight Press the control switch until the following screen appears After releasing the switch the display will momentarily show...

Страница 8: ... pull on the battery only Install a fresh battery and reinsert the battery in its compartment Reinstall the cover by gently pressing it into place while pulling out on the edges of the compartment and reinstall the four cover screws Take care to avoid overtightening the screws Always reuse the screws provided and do not damage or lose the o ring seal on each screw Cleaning The Ampstik can be clean...

Страница 9: ...r component FAIL CODE 3 Circuit failure FAIL CODE 4 Circuit failure The above fail codes may also trigger due to dirty pin connections causing errors in readings between the main board and molded housing assembly The unit will not enter into measurement mode if the self test has any of the above failures The unit will need to be returned to SensorLink for repair evaluation Repair SensorLink Servic...

Страница 10: ...Page 9 ...

Страница 11: ...ument 2 Subjected to misuse negligence or accident or 3 Connected installed adjusted or used otherwise than in accordance with the instructions furnished by SensorLink This Warranty is in lieu of any other warranty expressed or implied SensorLink reserves the right to make any changes in the design or construction of its instruments at any time without incurring any obligation to make any change w...

Страница 12: ...irements of that standard Serial Numbers_____________________________________________________ Model Numbers_____________________________________________________ I hereby certify that the True RMS Ammeter listed above has passed all tests defined in the SensorLink standard I also certify that I have reviewed the standard and test procedure and that they are sufficient in determining compliance with...