Section 2 SMT-200 Operations
Io
SENSOR
26
SMT-200 Operations and Technical Manual 2.00/R1
2.22.
2.9
Diagnostics
Runs the diagnostics and calibration routines
NOTE
: Use the Diagnostics test plug (part number SMT-200 C 0515) where indicated.
When exiting Diagnostics Test routine all settings are restored to original values.
To access this menu after selecting
Diagnostics
it you are required to enter
a
PIN
(Personal Identification Number) to access the menu. The
PIN
is the
serial number of the SMT-200 that can be found from the about menu.
Iopa
Is for calibration of the leakage test. To use this
connect the SMT-200 Diagnostics test plug or place
the required resistance (100M
Ω
default) between
the leak and geophone negative terminal. After
calibration the measured value should be about the
same as the expected leakage of the strings (or
100M
Ω
whichever is lower). On selection of this test
you are requested to enter the value for the test
resistor. Once this has been done the SMT-200 will
calculate the leakage current and thus be able to
accurately calculate leakage values.
Simulation
This will allow you to generate synthetic data
instead of real collected geophone data. This is
particularly useful for training and getting used to
the SMT-200 without needing to have a strings
connected. Make sure this is set to “off” before
testing actual strings. Test results are saved with
the comment SIMULATION.
Post Mortem
Shows how the unit shut down previously. The table
below shows the various messages and the
meaning of each.
Reset all
Restores all SMT-200 settings to their default
values and load the SM-4 10Hz 375
Ω
geophone
specification. Restored values are also saved to the
configuration file. (To reset display contrast press
the +/- key during booting).
Post Mortem messages
Normal Shut Down
Regular shut down with no problems detected
Temperature High
SMT-200 detected an internal temperature greater than 80°C
Temperature Low
SMT-200 detected an internal temperature lower than -40°C
Battery Low
SMT-200 detected a battery voltage less than 10.7 V
Timed Out
Shut down initiated by power saving function (set in Power menu)
Shut Down Error
Abnormal shut down (e.g. battery removed)
2>Setup menu...
9>Diagnostics...
Diagnostics menu
----------------
1>Iopa 0.07 nA
2 Simulation off
3 Post Mortem
4 Reset All
5 Display...
6 Hardware...
7 Tests...
0 Back to Setup
----------------
Calibrate Iopa
for Leakage test
Содержание SMT-200
Страница 2: ...Io SENSOR SMT 200 Operations and Technical Manual 2 00 R1 Page intentionally left blank ...
Страница 11: ...Io SENSOR Section 1 Introduction SMT 200 Operations and Technical Manual 2 00 R1 1 Section 1 Introduction ...
Страница 103: ...Io SENSOR Appendix B Test Overview SMT 200 Operations and Technical Manual 2 00 R1 93 Appendix B Test Overview ...
Страница 104: ...Appendix B Test Overview Io SENSOR 94 SMT 200 Operations and Technical Manual 2 00 R1 Page intentionally left blank ...
Страница 113: ...Io SENSOR Appendix C Parts List SMT 200 Operations and Technical Manual 2 00 R1 103 Appendix C Parts List ...
Страница 114: ...Appendix C Parts List Io SENSOR 104 SMT 200 Operations and Technical Manual 2 00 R1 Page intentionally left blank ...
Страница 126: ...Appendix D Schematics Layouts Io SENSOR 116 SMT 200 Operations and Technical Manual 2 00 R1 D 1 2 SMT 200 ADC ...
Страница 127: ...Io SENSOR Appendix D Schematics Layouts SMT 200 Operations and Technical Manual 2 00 R1 117 D 1 3 SMT 200 Filter ...
Страница 128: ...Appendix D Schematics Layouts Io SENSOR 118 SMT 200 Operations and Technical Manual 2 00 R1 D 1 4 SMT 200 Keyboard ...
Страница 129: ...Io SENSOR Appendix D Schematics Layouts SMT 200 Operations and Technical Manual 2 00 R1 119 D 1 5 SMT 200 DAC ...
Страница 131: ...Io SENSOR Appendix D Schematics Layouts SMT 200 Operations and Technical Manual 2 00 R1 121 D 1 7 SMT 200 Power ...
Страница 132: ...Appendix D Schematics Layouts Io SENSOR 122 SMT 200 Operations and Technical Manual 2 00 R1 D 1 8 SMT 200 Display ...
Страница 135: ...Io SENSOR Appendix D Schematics Layouts SMT 200 Operations and Technical Manual 2 00 R1 125 D 3 2 UC3906 B1 ...
Страница 136: ...Appendix D Schematics Layouts Io SENSOR 126 SMT 200 Operations and Technical Manual 2 00 R1 D 3 3 UC3906 B2 ...
Страница 139: ...Io SENSOR Appendix E Quick Reference SMT 200 Operations and Technical Manual 2 00 R1 129 Appendix E Quick Reference ...
Страница 140: ...Appendix E Quick Reference Io SENSOR 130 SMT 200 Operations and Technical Manual 2 00 R1 Page intentionally left blank ...
Страница 144: ...Appendix E Quick Reference Io SENSOR 134 SMT 200 Operations and Technical Manual 2 00 R1 Page intentionally left blank ...
Страница 145: ...Io SENSOR Appendix F Addendum SMT 200 Operations and Technical Manual 2 00 R1 135 Appendix F Addendum ...
Страница 146: ...Appendix F Addendum Io SENSOR 136 SMT 200 Operations and Technical Manual 2 00 R1 Page intentionally left blank ...
Страница 147: ...Io SENSOR Appendix F Addendum SMT 200 Operations and Technical Manual 2 00 R1 137 F 1 Addendum ...
Страница 148: ...Appendix F Addendum Io SENSOR 138 SMT 200 Operations and Technical Manual 2 00 R1 Page intentionally left blank ...
Страница 152: ...Notes Io SENSOR 142 SMT 200 Operations and Technical Manual 2 00 R1 Notes ...