Installation
1. Install battery in Main Unit
The 3 LEDs blink three times when
the battery is installed. After that, the
Main Unit is automatically set to
“service mode”.
Good to know:
5 minutes after the battery has been installed, the Main Unit automatically wakes
from “service mode” and starts the onboarding process. If the Main Unit is not
installed in the AED cabinet, it sends an error message to the Seluxit Cloud .
1
There is no cause for alarm in this case, as the Main Unit is reset during the
onboarding process which you will perform later, when the Main Unit is installed in
the AED cabinet.
2. Use the screws (including magnets)
to fasten the backplate to the Main
Unit. Magnets will thereby be placed
on all four corners.
NB: The QR code sticker (the monitoring unit’s unique
ID) is now on the Main Unit casing itself, and not on
the backplate as shown in the pictures.
1
Seluxit Cloud is called “Viasens by Seluxit: Secure IoT Infrastructure”
SLX Heartbeat for Zoll AED Plus, Installation Guide (version 1.4, September 2020)
6 of 15
Содержание SLX Heartbeat
Страница 1: ...Installation Guide...