background image

OWNER’S MANUAL

PRECISION MICROSCOPES

COLPOSCOPE

MODEL 935 LED

Содержание 935 LED

Страница 1: ...OWNER S MANUAL PRECISION MICROSCOPES COLPOSCOPE MODEL 935 LED...

Страница 2: ...2 Register your product warranty online http www seilermicro com warranty registration...

Страница 3: ...IONS FOR USE 4 PRESENTATION AND USE 4 COLPOSCOPE FLOOR STAND 4 THE COLPOSCOPE HEAD 5 CLEANING 5 HOW TO FOCUS THE COLPOSCOPE 6 OPTICAL AND FIELD MAGNIFICATION TABLE 7 TECHNICAL DATA 7 LED LIGHTSOURCE 8...

Страница 4: ...pes Read this Manual before setting up the equipment in order to avoid damage from improper usage For safe use of the colposcope it is necessary to be familiar with the details of its operation prior...

Страница 5: ...ly with soft and CIRCULAR movements Follow your laboratory s procedures to remove bloodstains or other contaminants from your equipment If the objective is too blotched change to another cotton swab f...

Страница 6: ...to the two pupils and the vision of the field appears as only one image With the inclined binoculars grasp each tube of the of the binoculars and move the tubes in or out so that as it was already sai...

Страница 7: ...Light Guide Filter Green Focusing Fine Focus 11mm B Colposcope Stand Angular Rotation 360 Degrees Distance between the 1016mm 1193mm floor and the objective Base Five casters two with brakes Total Wei...

Страница 8: ...or nitrate oxide To prevent any potential electro magnetic interference do not use any kind of cellular phone near the light source This product should be used only with type BF endoscopic instrument...

Страница 9: ...ation Continuous operation Water Resistant Not Protected Equipment IPX0 Operating Environment Temperature Relative Humidity Air Pressure 0 to 40 C 32 to 104 F 0 to 85 rh non condensing 700 to 1060 hPa...

Страница 10: ...tion 1 AC main inlet Accepts AC power cord 2 Fuse 2A 250V 1 4 x 1 time delay 3 Grounding Connector For potential equalization 4 Venting grids Allows adequate ventilation and cooling of the light sourc...

Страница 11: ...is equipment may be subjected to radio interference To avoid adverse electromagnetic effects DO NOT operate this equipment near RF energy equipment 5 2 CONNECTING THE FIBER OPTIC CABLE Turn intensity...

Страница 12: ...ty from the date of shipment on workmanship and all defects of materi al excluding replaceable sweatbands headband fiberoptic cable and lamps Should your product prove to have such defects within one...

Страница 13: ...be found at http www eiae org Please contact customer service to issue a return authorization to return product to manufacturer at the end of product life 11 TROUBLESHOOTING Problem Solution The power...

Страница 14: ...14 Model Number Serial Number Date Purchased Location Purchased OWNER S RECORD NOTES...

Страница 15: ...68 3601 E mail micro seilerinst com www seilerinst com Microscope Division Manager Dane Carlson ext 365 dcarlson seilerinst com Corporate Marketing Sales Representative Nicole Rasch ext 345 nrasch sei...

Страница 16: ...PRECISION MICROSCOPES 3433 Tree Court Industrial Blvd St Louis Missouri 63122 Toll Free 800 489 2282 Local 314 968 2282 Email micro seilerinst com Web www seilerinst com Version 09 22 14...

Отзывы: