SPX100 SENSOR
DUST EMISSIONS CONTROLLER
Notice Technique
FI 72.0110.0316E
Page 1 sur 27
SPX 100 V4
TECHNICAL NOTICE
Страница 1: ...SPX100 SENSOR DUST EMISSIONS CONTROLLER Notice Technique FI 72 0110 0316E Page 1 sur 27 SPX 100 V4 TECHNICAL NOTICE ...
Страница 2: ...5 Date and time adjustment 6 Setting the unit 7 Setting the display format 8 View measuring scale 9 Setting the 4 20 mA scale 10 Alarm average duration adjustment 11 Auto sensitivity 12 Auto zero 13 Average calculation 14 Calibration 15 Language adjustment 16 Alarm adjustment Alarm reading VI INSTALLATION START page 22 VII CALIBRATION PRINCIPLE page 22 VIII General instructions for safety assembly...
Страница 3: ...ts extremity is projected into the particle flow monitored the signal The control unit receives the data from the sensor averaging measure converses with the user display keys automatic zero calibration stores and dates the datas 1 month of recordings commands the alarm an output 4 20 mA in option of the trend The SPX 100 can be connected with a PC line RS232 and software in option directly or thr...
Страница 4: ...SPX100 SENSOR DUST EMISSIONS CONTROLLER Notice Technique FI 72 0110 0316E Page 4 sur 27 PRINCIPLE PLAN ...
Страница 5: ... flow at the sensor 85 C standard sensor 250 C sensor with thermal separator 250 C please contact us Storage temperature 20 to 70 C Alarm threshold Adjusting from 0 to full scale shown in alarm adjustment Averaging measure From 1 to 60 minutes Calendar Retention 11 years without supply Accuracy 1 minute per month maxi Leap year year 2000 1 month memory Save by battery kind R6 Retention 10 years wi...
Страница 6: ... or the duct to be monitored Filetage 1 BSP 26 34 GAS Placed on a rectilinear position 5 diameters before 2 diameters after mini N B If stack 5 diameters after grounded metallic duct 5 meters before the sensor mini The continuity of earth between the sensor and the section of casing must be ensured The sensor length must be adjusted to the duct ...
Страница 7: ...tice Technique FI 72 0110 0316E Page 7 sur 27 2 Base To access the terminal blocks put off the connections cover NOTE The relay contacts are drawing main supply off When the supply voltage comes on the relay pulls up if no alarm FUSE PROTECTION ...
Страница 8: ... 0316E Page 8 sur 27 SUPPLY CHOICE To access this put off the front panel and the upper board BE CAREFUL The SPX 100 leaves our workshops for a 230V supply voltage WARNING If you change the supply voltage you must update the data plate located on the outside of the box ...
Страница 9: ...the last 1 month datas scrolling memory 1 storing each minutes To calculate the average value of the dust concentration on 1 month maxi see average calculation on 24 hours Possibility to read the datas via a PC Incident analysis graphics outputs Report outputs etc Display of the instantaneous and averaged trend from 1 to 60 mm in mg m3 µg m3 or of full scale see measure reading adjustment If the m...
Страница 10: ...d keep it pressed Press the valid key see alarm adjustment option of a non maintained alarm please consult us 3 4 20 mA in option The 4 20 mA output scale is adjustable The 4 20 mA output recopies the instantaneous measure or the average measure according to the display see measure reading adjustment 4 mA for 0 measure 20 mA for the value of the 4 20 mA input scale 4 Connection Each time the base ...
Страница 11: ...EMISSIONS CONTROLLER Notice Technique FI 72 0110 0316E Page 11 sur 27 V ADJUSTMENTS All the adjustments are made with an alphanumerical display a bargraph and some keys in front panel 1 Front panel front panel description ...
Страница 12: ...g the VALID key The keys that have a function for the selected parameter adjustment are symbolised on the display up key down key left key right key valid key Passing to the next step with the VALID key valid the previous parameter All the parameters are stored into memory Adjustment rolling menu NOTE automatic return to the read measurement position occurs after 1 minute of inactivity ...
Страница 13: ...lied duration average duration If the alarm is cancelled when alarm activated the averaged measure is initialised with the instantaneous measure in progress So wait the average duration for a new averaged measure NOTE The 4 20 mA output recopies the displayed measure instantaneous or averaged according to the selected choice see above see running 4 20 mA 4 Password adjustment or CODE The password ...
Страница 14: ...is pressed if the password is good the display shows To change the password press the right key otherwise the valid key To adjust a new password select the digit you want to adjust with the left and right key and press the up and down keys to increase or decrease the selected digit Then press the valid key the new password is activated BE CAREFUL If you lose the password all the adjustments stay f...
Страница 15: ... good select the value you want to adjust with the left and right key and press the up and down key to increase or decrease the selected value 6 Unit adjustment Set the display units with the up and down buttons mg m3 by default or µg m3 or g m3 or 7 Display format adjustment Set the display format with the up and down buttons X XXXX or XX XXX or XXX XX by default or XXXX X or XXXXX To be adjusted...
Страница 16: ...onfirm this WARNING if the 4 20 mA scale is greater than the value of the measuring scale 20 mA will never be obtained at the output 10 Alarm average duration adjustment Adjustment of the average duration 1 2 5 10 15 20 25 30 35 40 45 50 55 ou 60 minutes It s this average value calculated on this duration who s be read in reading measure see measure reading adjustment It s with this average value ...
Страница 17: ...execution by pressing the right key else the VALID key allows quitting this function c Enter the required value of the measuring scale This must be selected as a function of an alarm limit which is to be assigned and as a function of the current emissions value Where possible the aim is to have a scale value corresponding to 10 times the current emissions level e g value of the current estimated e...
Страница 18: ...ect a value which covers representative operation of the installation unclogging arrival of unstable material etc f The SPX 100 makes some tests to obtain an optimal sensor sensibility The display shows a message like that g At the end of the test 2 cases The procedure finishes correctly the device returns to measurement view with the new settings The procedure does not finish correctly the displa...
Страница 19: ...ro The auto zero allows the offset adjustment of the device according with its environment where it is fitted noise electronic offset When we arrive in this step the SPX 100 asks If the VALID key is pressed next step without doing an auto zero If the right key is pressed beginning of the procedure auto zero It s automatic and lasting 75 seconds Wait the end of the procedure NOTE Unnecessary to put...
Страница 20: ... 100 a Press the right key b Enter the date time of the start of the period using the up down right and left buttons c Enter the date time of the end of the period in the same way d The average emissions level is calculated and then displayed NOTE If the result is flicking it is because one data on the period selected at least isn t up dated This may mean that the device was switched off during th...
Страница 21: ...as in the average calculation section b The result obtained by the SPX 100 is calculated and then displayed Press the VALID button c Correct the value displayed with your result using the up down right and left buttons then press the VALID button NOTE the calibration of the SPX 100 affects the measuring scale when the calibration has been done the SPX 100 adjusts its calculations accordingly BE CA...
Страница 22: ... the averaged measure is initialised with the instantaneous measure So if the instantaneous measure is upper than the alarm threshold the alarm is activated again The alarm is stored into memory even if the supply drops and it is keeping maintained even if the measure drops below the alarm threshold For operation with a non maintained alarm please consult us BE CAREFUL The measure full scale is wi...
Страница 23: ... display units and format as a function of the required measuring scale Go to auto sensibility and valid VII calibration principle Recalibrating an SPX 100 from a weighted reference measurement On the day of the measurement the SPX 100 must be operating in parallel with the mass measurement Several days later once the results of the reference measurement have been obtained perform the calibration ...
Страница 24: ... between phases and not between phase and neutral so as to avoid any accidental over voltage caused when the neutral is cut before the phases In the event of a problem please contact Sefram COMMISSIONING AND USAGE Commissioning is authorised only after duly establishing that the device the machine or the installation in which the device has been integrated in a compliant manner satisfies all of th...
Страница 25: ...is forbidden You must send us back the device in our workshops otherwise the certification and the guarantee are lost In case of failure no action is permitted and the unit must be returned to the following address X CERTIFICATE The SPX100 base and sensor respects the European directives CEM BT and ATEX which concerns it However it must be used correctly in applications for which it is intended an...
Страница 26: ...SPX100 SENSOR DUST EMISSIONS CONTROLLER Notice Technique FI 72 0110 0316E Page 26 sur 27 XI PHYSICAL DIMENSIONS 1 Sensor STANDARD WITH THERMAL SEPARATOR ...
Страница 27: ...SPX100 SENSOR DUST EMISSIONS CONTROLLER Notice Technique FI 72 0110 0316E Page 27 sur 27 2 Control unit SEFRAM RESERVES THE RIGHT TO MAKE CHANGE TO ITS PRODUCTS ...