26
Technical characteristics
EcoStar Info, EcoStar Auto
With GoodKnight H
2
O
humidifier
Test pressure (cmH
2
O)
4
8
12
16
20
Respiratory rate
Resp./min
Respiratory rate
Resp./min
Respiratory rate
Resp./min
Respiratory rate
Resp./min
Respiratory rate
Resp./min
10
15
20
10
15
20
10
15
20
10
15
20
10
15
20
Tubing 22 mm diameter:
Delta relative to highest
pressure (cmH
2
O)
0.13 0.14 0.21 0.35 0.37 0.40 0.38 0.38 0.47 0.35 0.35 0.46 0.29 0.34 0.44
Delta relative to lowest
pressure (cmH
2
O)
0.13 0.21 0.27 0.05 0.14 0.25 0.12 0.21 0.26 0.27 0.35 0.41 0.43 0.50 0.56
Delta relative to dynamic
pressure (cmH
2
O)
0.04 0.06 0.06 0.08 0.15 0.13 0.10 0.16 0.20 0.15 0.23 0.25 0.16 0.29 0.33
Tubing 15 mm diameter:
Delta relative to highest
pressure (cmH
2
O)
0.09 0.16 0.32 0.34 0.43 0.63 0.36 0.52 0.71 0.32 0.51 0.75 0.31 0.52 0.77
Delta relative to lowest
pressure (cmH
2
O)
0.26 0.31 0.35 0.17 0.23 0.26 0.25 0.25 0.32 0.40 0.40 0.48 0.53 0.56 0.61
Delta relative to dynamic
pressure (cmH
2
O)
0.06 0.08 0.08 0.13 0.14 0.15 0.16 0.22 0.23 0.22 0.24 0.33 0.23 0.35 0.31
Tubing 15 mm diameter and high-efficiency filter:
Delta relative to highest
pressure (cmH
2
O)
0.11 0.18 0.43 0.22 0.30 0.52 0.23 0.31 0.49 0.17 0.31 0.42 0.12 0.26 0.36
Delta relative to lowest
pressure (cmH
2
O)
0.32 0.42 0.53 0.28 0.36 0.47 0.42 0.50 0.60 0.61 0.64 0.81 0.79 0.83 0.98
Delta relative to dynamic
pressure (cmH
2
O)
0.07 0.10 0.09 0.11 0.15 0.13 0.15 0.19 0.24 0.20 0.26 0.26 0.30 0.32 0.31
Maximum flow and pressure
(According to ISO 17510-1:2009. Appendix CC.1 and ISO 80601-2-70:2014 §201.12.1.103)
Test pressure (cmH2O)
4
8
12
16
20
Maximum flow (l/min) provoking
a pressure drop of 1 cmH
2
O
at the
opening of the patient side
connection
Tubing 22 mm diameter
>130
180
180
170
150
Pressure measurement (cmH
2
O)
at 40 l/min at the patient
interface
Tubing 22 mm diameter
3.9
8.1
12.1
16.0
19.9
Sound pressure and sound power levels
(according to ISO 17510-1:2009 and et ISO 80601-2-70:2014 §201.9.6.2.1.101)
Without humidifier
With GoodKnight H
2
O humidifier
and tubing 15 mm diameter
Sound pressure level
27.5 dB(A)
29.5 dB(A)
Sound power level
35.5 dB(A)
37.5 dB(A)