User Guide
©2008
-2021 Seeed Technology Co., Ltd. All rights reserved. solution.seeedstudio.com
13
/
22
PROTOCOL
0x0202 /514
UINT16
R/W
3/6/16
0-1
0:Modbus RTU
1:Modbus ASCii
0:Modbus
RTU
PARITY
0x0203 /515
UINT16
R/W
3/6/16
0-2
0:None
1:Even
2:Odd
0:None
Parity
DATABITS
0x0204 /516
UINT16
R/W
3/6/16
1
1:8 databits
1:8 databits
STOPBITS
0x0205 /517
UINT16
R/W
3/6/16
0-1
0:1 stopbit
1:2 stopbits
0:1 stopbit
RESPONSEDELAY
0x0206 /518
UINT16
R/W
3/6/16
0-255 for 0-2550
milliseconds
0
ACTIVEOUTPUTIN
TERVAL
0x0207 /519
UINT16
R/W
3/6/16
0-255 for 0-255 seconds.
0
NOTE: UINT16:16 bit unsigned integer, INT16:16bit signed integer
NOTE: RO: Register is Read Only, R/W: Register is Read/Write
NOTE: HEX is Hexadecimal (data with 0x/0X prefix), DEC is Decimal
7.3 Modbus Register Detail Descripton
TEMPERATURE
Data Range
-4000-8000 For -40.00~80.00°C
Default: N/A
Power Down Save
N/A
Note:Temperature value (Binary complement).
Example:When REGISTER = 0x0702 (HEX format), then
VALUE=(0x07*256+0x02)/100=17.94°C.When REGISTER=FF05H (HEX format),then
VALUE=((0xFF*256+0x05)-0xFFFF-0x01)/100 =(0xFF05-0xFFFF-0x01)/100=-2.51°C.
EC ---Electrical Conductivity
Data Range
0-20000 For 0-20000us/cm
Default: N/A
Power Down Save
N/A
Note:Electrical Conductivity.
Example:When REGISTER = 0x0702 (HEX format), then VALUE=(0x07*256+0x02)=1794us/cm
SALINITY--- Salinity