HF Series Generators -- RAD Console
Operation
OM-0279R1
34
Table 3-2 (cont.)
Error Codes
ERROR
DESCRIPTION
WHAT TO DO
E25
Battery Fault. The batteries charge level is momentarily
low, or some batteries are discharged or damaged.
(Only in
Generators working with batteries).
Press the “
AEC Reset
” push-button.
Wait 5 minutes before making a new exposure.
If the error code persists, turn the Generator OFF and ON.
If the equipment remains inoperative, turn it OFF and call Field Service.
E26, E27
System failure.
Press the “
AEC Reset
” push-button.
If the error code persists, turn the Generator OFF and ON.
If the equipment remains inoperative, turn it OFF and call Field Service.
E33
Serial Communication error.
Press the “
AEC Reset
” push-button.
Check that communication cable between Generator and console is
properly connected.
If the error code persists, turn the Generator OFF and ON.
If the equipment remains inoperative, turn it OFF and call Field Service.
E34
Technique error. If it activates during exposure it means
that the exposure has been interrupted by the “
Security
Timer
” because of a system failure. Call Field Service.
This error can also be shown:
-- after an APR technique selection to advise that exposure
parameters displayed on the console are not the values
stored for this APR technique. Exposure parameters are
adapted by the Generator to another enable values.
-- after the “ABC” push-button selection, when ABC is not
enable.
E35
Door Open error. The X-ray room door is open when the
X-ray equipment is in use.
These errors do not require to press the “
AEC Reset
” push-button, theirs
i dicatio s disappea a to aticall
E36
Heat Units error. The X-ray Tube thermostat / pressurestat
is open due to the tube housing is overheated (housing is
too hot, wait for the housing to cool) or a thermostat /
pressurestat mal-function (housing is cool). Heat units may
raise to any value.
indications disappear automatically.
If the error code persists, turn the Generator OFF and ON.
If the equipment remains inoperative, turn it OFF and call Field Service.
E37
Tube Overload error. The technique selected is beyond the
X-ray tube ratings or present conditions of the X-ray tube
inhibit the exposure (anode overheated). Parameters for
next exposure may be temporally limited by the Generator
(change the exposure values or wait for the X-ray tube to
cool). Check that heat units available are lower than the
calculated for the next exposure (heat units close to zero).
Reduce exposure factors or wait for the X-ray tube to cool.
E41 to E46
System failure related to Dosimeter.
Press the “
AEC Reset
” push-button.
If the error code persists, turn the Generator OFF and ON.
If the equipment remains inoperative, turn it OFF and call Field Service.
E47
Capacitors are not charged when “
Prep
” control is activated.
The exposure is inhibited until the Capacitors are charged.
Press the “
AEC Reset
” push-button.
Wait one minute for Capacitor charging before activating “
Prep
” control.
If the error code persists, turn the Generator OFF and ON.
If the equipment remains inoperative, turn it OFF and call Field Service.
E48
Collimator Error. A failure on the Automatic Collimator has
been detected (blades are full open or in movement during
exposure, etc. )
Press the “
AEC Reset
” push-button.
If the error code persists turn the Generator OFF and ON
E50
Exposure has been aborted by the Operator.
If the error code persists, turn the Generator OFF and ON.
If the equipment remains inoperative, turn it OFF and call Field Service.
E51 to E93
System failure related to High Speed Rotor Controller.
q p
p
,
E95
Exposure aborted by the AEC Rapid Termination.
Press the “
AEC Reset
” push-button. Select the correct Ion Chamber or
modify parameters. Repeat the exposure.
If the equipment remains inoperative, turn it OFF and call Field Service.
E96, E97
System failure related to Capacitor charge (only for
Capacitor Powered Generator).
Press the “
AEC Reset
” push-button.
If the error code persists, turn the Generator OFF and ON.
If the equipment remains inoperative, turn it OFF and call Field Service.
E98
Service Mode Active.
Press the “
AEC Reset
” push-button and call Field Service. This error
does not inhibit normal operation.