HF Series Generators -- RAD Console
Operation
OM-0279R1
16
3.1
RADIOGRAPHY AND GENERAL CONTROLS
3.1.1
POWER ON / OFF
ON:
The Generator is turned ON by pressing this push-button. This starts a
power-up routine which is shown on the Console (a.e. P05.3 = Vers.5, Rev.3).
After the power-up routine the last workstation used will be automatically
selected.
OFF
: The Generator is turned OFF by pressing this push-button.
For Generators equipped with “High Speed Rotor Controller”, if the X-ray Tube
is rotating when pressing the “
OFF
” push-button, the Generator will stop the
anode immediately and then the unit will be turned off (approx. 3 seconds). The
equipment only turns off if “
Preparation
” is not activated.
IN THE EVENT OF AN EMERGENCY, ISOLATE THE X-RAY
GENERATOR FROM THE MAINS POWER BY FORCIBLY
DEPRESSING THE X-RAY ROOM “EMERGENCY OFF
SWITCH” (USUALLY A RED MUSHROOM-SHAPED SWITCH).
THIS SWITCH SHOULD BE LOCATED ON OR NEAR TO THE
X-RAY ROOM MAINS POWER PANEL, USUALLY PLACED
NEAR TO THE GENERATOR CONTROL CONSOLE. MORE
THAN ONE OF THESE SWITCHES MAY BE PLACED AROUND
THE ROOM FOR GREATER ACCESSIBILITY.
3.1.2
WORKSTATIONS SELECTION
The workstations are configured according to the customer preferences during
the installation procedure. The following push-buttons select their respective
workstation (illuminated push-buttons).
If a push-button is not configured during the installation, the respective
push-button(s) can not be selected during operation.