HF Series Generators -- RAD Console
Operation
OM-0279R1
10
Do not use devices which intentionally transmit RF Signals
(Cellular Phones, Transceivers or Radio Controlled
Products) in the vicinity of this equipment as it may cause
performance outside the published specifications. Keep the
power to these type devices turned off when near this
equipment. Is the responsibility of the operator to instruct
patients and other people who may be around this equipment
to fully comply with the above requirement.
The manufacturer is not responsible for any interference
caused by using other than recommended interconnect
cables or by unauthorized changes or modifications to this
equipment.
GUIDANCE AND MANUFACTURER’S DECLARATION -- ELECTROMAGNETIC EMISSIONS
This X--Ray Generator is intended for use in the electromagnetic environment specified below.
The customer or the user of this X--Ray Generator should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment -- guidance
RF emissions
CISPR 11
Group 1
This X--Ray Generator uses RF energy only for
its internal function. Therefore, its RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions
CISPR 11
Class A
This X--Ray Generator is suitable for use in all
Harmonic emissions
IEC 61000--3--2
Class A
This X--Ray Generator is suitable for use in all
establishments other than domestic and those
directly connected to the public low--voltage
power supply network that supplies buildings
Voltage fluctuations/flicker emissions
IEC 61000--3--3
Complies
power supply network that supplies buildings
used for domestic purposes.