Modifying presets
NOTE
Your settings apply to the current measurement procedure. If you
stop the measurement procedure
, the presets defined by the administrator become active again.
1. Ensure that no SpO
2
sensor has been applied to the patient.
2. Press the
SpO
2
field.
The
SpO
2
settings
dialog window opens.
The presets are displayed (in this case: Masimo SET
®
pulse oximetry).
3. Press the desired sensitivity:
4. Press the
Confirm
button.
The dialog window closes.
The changed settings are used for the current measurement.
5. Start the SpO
2
measurement as described in the section entitled
Weight and height
The device provides several options for adding the parameters weight and
height. Your administrator specifies which of the options you can use. Further
information is available here:
.
Masimo SET
®
SpO
2
module
Mode
Indication
Normal
• Normal perfusion
• Mild perfusion disorders
Adaptive Probe Off Detection
(APOD)
Vigorous patient movements
Maximum
• Poor perfusion
• Severely disrupted signal, for
example due to indoor lighting or
direct sunlight
seca SpO
2
module
Mode
Motion tolerance
Stable
High
Normal
Normal
Sensitive
Low