Managing system components • 161
17
-1
0-
05
-3
53
-0
02
g_
10
-2
02
0
B
4.5 Retrofitting the measuring mat (
seca mVSA 535
only)
NOTICE!
Malfunction
If your device is connected to an EMR system, the Measure
bioimpedance function can
not
be used.
►
Do not retrofit a measuring mat.
►
Ensure that the
Measuring mat active
checkbox is deactivated.
seca mVSA 535
can be retrofitted with the
seca mBCA 531
measuring mat
to perform bioimpedance measurements
Optional accessories and spare
1. Start up the measuring mat as described in the corresponding assembly
instructions.
2. Activate the measuring mat in the device:
a) Press the
System
tab
b) Press the
Measuring mat
button
c) Activate the
Measuring mat active
checkbox
3. Follow the on-screen instructions.