Technical data • 123
17
-1
0-
05
-3
53
-0
02
g_
10
-2
02
0
B
Clinical studies
Clinical studies form the scientific basis for analyzing body composition using
this device. The study results are integrated in the device software in the form
of reference values.
For some analysis parameters, the reference used depends on the patient's
ethnicity. The device automatically uses ethnicity-dependent references to
suit the corresponding entry in the seca patient file
Details about the clinical studies can be found on our website,
www.seca.com.
Accuracy of predictive formulas
Accuracy
(8-point bioimpedance measurement, frequencies 5 kHz and 50 kHz
Segments: Right half of body, left half of body)
• Impedance (at phase angle 0°)
• Phase angle (at phase angle 0°, impedance 200 Ω to 1000 Ω)
±5 Ω
0.5°
Measuring method
Standard deviation (SEE) for predictive formulas in this device
a
Ethnicity:
Caucasian
Afro-American
Asian
South and Central
American
Other
Parameter
SEE
SEE
SEE
SEE
SEE
FFM
2.50 kg
2.21 kg
2.54 kg
2.62 kg
2.49 kg
TBW
2.0 l
1.8 l
1.4 l
1.4 l
1.7 l
ECW
1.1 l
0.9 l
0.9 l
0.7 l
0.9 l
SMM
left arm
0.19 kg
0.28 kg
0.19 kg
0.16 kg
0.21 kg
SMM
right arm
0.22 kg
0.30 kg
0.21 kg
0.17 kg
0.23 kg
SMM
left leg
0.81 kg
0.71 kg
0.81 kg
0.83 kg
0.79 kg
SMM
right leg
0.68 kg
0.66 kg
0.76 kg
0.71 kg
0.70 kg
SMM
total
1.8 kg
2.0 kg
1.7 kg
1.7 kg
1.8 kg
VAT
0.5 l
0.6 l
0.6 l
1.2 l
0.8 l
a. In the USA, a study was conducted on 130 healthy adults of different ethnicities. The aim of the study was to validate the para
-
meters determined using seca formulas against clinically established reference methods. The results of this comparative study are
shown in the table above. The table shows the standard deviation (SEE) by ethnicity for the parameters determined using seca's
in-house formulas.