Section 4: Setting Up MicroTSG
SBE 45
27
Command Descriptions
This section describes commands and provides sample outputs.
See
Appendix III: Command Summary
for a summarized command list.
When entering commands:
•
Input commands to the MicroTSG in upper or lower case letters and
register commands by pressing the Enter key.
•
The MicroTSG sends
? CMD
if an invalid command is entered.
•
If the system does not return an
S>
prompt after executing a command,
press the Enter key to get the
S>
prompt.
•
If in quiescent state, re-establish communications by clicking Connect on
the Toolbar or pressing the Enter key to get an
S>
prompt.
Status
Command
DS
Display operating status and setup parameters.
Equivalent to Status on Toolbar.
List below includes, where applicable,
command used to modify parameter.
•
Firmware version and serial number
•
Logging status -
logging data, not
logging data,
or
unknown status
•
Sample interval time [
Interval=
]
•
Output conductivity with each sample
[
OutputCond=
]?
•
Output salinity with each sample
[
OutputSal=
]?
•
Output sound velocity with each sample
[
OutputSV=
]? If yes, sound velocity
algorithm [
SVAlgorithm=
].
•
Start sampling automatically when power on
[
AutoRun=
]?
•
Go to sleep after taking single sample
[
SingleSample=
]?
•
Go to sleep after 2 minutes of inactivity
[
AutoOff=
]?
•
A/D cycles to average per sample
[
NCycles=
]
•
Output format (does not appear if
OutputFormat=0
. Shows
conductivity
leading space is suppressed
if
OutputFormat=1
; shows
conductivity and
salinity order reversed
if
OutputFormat=2
)
Example:
(user input in bold)
S>
DS
SBE45 V 1.1b SERIAL NO. 1258
not logging data
sample interval = 10 seconds
[
Interval=
]
output conductivity with each sample
[
OutputCond=
]
do not output salinity with each sample
[
OutputSal=
]
do not output sound velocity with each sample
[
OutputSV=
and
SVAlgorithm=
]
do not start sampling when power on
[
AutoRun=
]
do not power off after taking a single sample
[
SingleSample=
]
do not power off after two minutes of inactivity
[
AutoOff=
]
A/D cycles to average = 4
[
NCycles=
]
32
Содержание SBE 45 MicroTSG
Страница 2: ...This page intentionally left blank ...
Страница 4: ...Manual Generation Date 13 February 2012 1 ...
Страница 6: ...This page intentionally left blank 3 ...
Страница 7: ...4 ...
Страница 98: ...CALIBRATION SHEETS SBE 45 Temperature Calibration S N 0402 1 SBE 45 Conductivity Calibration S N 0402 2 95 ...
Страница 112: ...109 ...
Страница 118: ...6 Appendix I Example Conductivity Calibration Sheet 115 ...
Страница 119: ...7 Appendix II Example Temperature Calibration Sheet 116 ...
Страница 129: ...126 ...
Страница 130: ...127 ...
Страница 131: ...128 ...
Страница 132: ...129 ...
Страница 133: ...130 ...
Страница 134: ...131 ...
Страница 135: ...132 ...
Страница 136: ...133 ...
Страница 137: ...134 ...
Страница 138: ...135 ...