APPLICATION NOTE NO. 2D
Revised October 2010
Instructions for Care and Cleaning of Conductivity Cells
This application note presents new recommendations (as of October 2006), based on our recent research, for cleaning and
storing conductivity sensors. In the past, Sea-Bird had recommended cleaning and storing conductivity sensors with a Triton
X-100 solution, and cleaning conductivity sensors with an acid solution.
Our latest research leads us to recommend adding
the use of a dilute bleach solution to eliminate growth of bio-organisms, and eliminating the use of acid in most cases
.
The application note is divided into three sections:
General discussion
Rinsing, cleaning, and storage procedures
Cleaning materials
General Discussion
Since any conductivity sensor’s output reading is proportional to its cell dimensions, it is important to keep the cell clean of
internal coatings. Also, cell electrodes contaminated with oil, biological growths, or other foreign material will cause low
conductivity readings. A desire to provide better control of growth of bio-organisms in the conductivity cell led us to develop
revised rinsing and cleaning recommendations.
A dilute bleach solution is extremely effective in controlling the growth of bio-organisms in the conductivity cell. Lab
testing at Sea-Bird indicates no damaging effect from use of a dilute bleach solution in cleaning the conductivity cell. Sea-
Bird now recommends cleaning the conductivity sensor in a bleach solution.
Triton X-100 is a mild, non-ionic surfactant (detergent), valuable for removal of surface and airborne oil ingested into the
CTD plumbing as the CTD is removed from the water and brought on deck. Sea-Bird had previously recommended, and
continues to recommend, rinsing and cleaning the conductivity sensor in a Triton solution.
Sea-Bird had previously recommended acid cleaning for eliminating bio-organisms or mineral deposits on the inside of
the cell. However, bleach cleaning has proven to be effective in eliminating growth of bio-organisms; bleach is much
easier to use and to dispose of than acid. Furthermore, data from many years of use shows that mineral deposits are an
unusual occurrence. Therefore, Sea-Bird now recommends that, in most cases, acid should not be used to clean the
conductivity sensor.
In rare instances
, acid cleaning may still be required for mineral contamination of the
conductivity cell.
Sea-Bird recommends that you return the equipment to the factory for this cleaning if it is
necessary
.
Sea-Bird had previously recommended storing the conductivity cell filled with water to keep the cell wetted, unless the cell
was in an environment where freezing is a possibility (the cell could break if the water freezes). However, no adverse affects
have been observed as a result of dry storage, if the cell is rinsed with fresh, clean water before storage to remove any salt
crystals. This leads to the following revised conductivity cell storage recommendations:
Short term storage (less than 1 day, typically between casts): If there is no danger of freezing, store the conductivity cell
with a dilute bleach solution in Tygon tubing looped around the cell. If there is danger of freezing, store the conductivity
cell dry, with Tygon tubing looped around the cell.
Long term storage (longer than 1 day): Since conditions of transport and long term storage are not always under the
control of the user, we now recommend storing the conductivity cell dry, with Tygon tubing looped around the cell ends.
Dry storage eliminates the possibility of damage due to unforeseen freezing, as well as the possibility of bio-organism
growth inside the cell. Filling the cell with a Triton X-100 solution for 1 hour before deployment will
rewet
the cell
adequately.
Note that the Tygon tubing looped around the ends of the conductivity cell, whether dry or filled with a bleach or Triton
solution, has the added benefit of keeping air-borne contaminants (abundant on most ships) from entering the cell.
Sea-Bird Electronics, Inc.
13431 NE 20
th
Street
Bellevue, WA 98005
USA
Phone: (425) 643-9866
Fax: (425) 643-9954
E-mail: [email protected]
Web: www.seabird.com
99
Содержание SBE 45 MicroTSG
Страница 2: ...This page intentionally left blank ...
Страница 4: ...Manual Generation Date 13 February 2012 1 ...
Страница 6: ...This page intentionally left blank 3 ...
Страница 7: ...4 ...
Страница 98: ...CALIBRATION SHEETS SBE 45 Temperature Calibration S N 0402 1 SBE 45 Conductivity Calibration S N 0402 2 95 ...
Страница 112: ...109 ...
Страница 118: ...6 Appendix I Example Conductivity Calibration Sheet 115 ...
Страница 119: ...7 Appendix II Example Temperature Calibration Sheet 116 ...
Страница 129: ...126 ...
Страница 130: ...127 ...
Страница 131: ...128 ...
Страница 132: ...129 ...
Страница 133: ...130 ...
Страница 134: ...131 ...
Страница 135: ...132 ...
Страница 136: ...133 ...
Страница 137: ...134 ...
Страница 138: ...135 ...