Rotary Evaporator
15
remove the foreign object if any.
•
ERR 002, indicating abnormal speed.
-
Turn off the instrument and restart it.
•
ERR 003, indicating excessively high temperature.
-
Turn off the heating bath and restart it when its
temperature falls to the room temperature.
•
ERR 004 refers to the sensor failure
-
Change of temperature sensor
If the above faults can not be cleared, please contact the
manufacturer/supplier.
7.
Maintenance and cleaning
Operate and maintain the product properly, so that it is in a
good working state, which can extend the service life of the
product. In routine service, keep the product dry and clean,
remove the spilled liquid quickly, clean the outer surface
with a non-grinding cleaner, and do not connect the power
supply until all surfaces are dry. If liquid or moist solid
enters the product, please disconnect the power supply
quickly and leave off, and contact the manufacturer /
supplier for more advice.
•
Keep the product clean, and the cleaning solution is
not allowed to flow into the machine.
•
Power must be disconnected before maintenance and
cleaning, and please use our recommended methods to
clean the product. The method to clean:
Dye
Isopropanol
Building
materials
Aqueous solution / isopropanol with
active agent
Cosmetic
Aqueous solution / isopropanol with
active agent
Food
Aqueous solution with active agent
Fuel oil
Aqueous solution with active agent
•
You can consult the manufacturer about the materials
that are not listed in the above table. Before using
other cleaning methods, the user must confirm with the
manufacturer / supplier that the method will not
damage the instrument. When cleaning the product,
please wear suitable protective gloves.