background image

Rotary Evaporator 

 

15 

remove the foreign object if any. 

 

ERR 002, indicating abnormal speed. 

-

 

Turn off the instrument and restart it. 

 

ERR 003, indicating excessively high temperature. 

-

 

Turn  off  the  heating  bath  and  restart  it  when  its 

temperature falls to the room temperature.   

 

ERR 004 refers to the sensor failure 

-

 

Change of temperature sensor 

If  the  above  faults  can  not  be  cleared,  please  contact  the 

manufacturer/supplier. 

7.

 

Maintenance and cleaning   

Operate and maintain the product properly, so that it is in a 

good working state, which can extend the service life of the 

product. In routine service, keep the product dry and clean, 

remove  the  spilled  liquid  quickly,  clean  the  outer  surface 

with a non-grinding cleaner, and do not connect the power 

supply  until  all  surfaces  are  dry.  If  liquid  or  moist  solid 

enters  the  product,  please  disconnect  the  power  supply 

quickly  and  leave  off,  and  contact  the  manufacturer  / 

supplier for more advice. 

 

Keep  the  product  clean,  and  the  cleaning  solution  is 

not allowed to flow into the machine. 

 

Power  must  be  disconnected  before  maintenance  and 

cleaning, and please use our recommended methods to 

clean the product. The method to clean: 

Dye 

Isopropanol 

Building 

materials 

Aqueous solution / isopropanol with 

active agent 

Cosmetic 

Aqueous solution / isopropanol with 

active agent 

Food 

Aqueous solution with active agent 

Fuel oil 

Aqueous solution with active agent 

 

You  can  consult  the  manufacturer  about  the  materials 

that  are  not  listed  in  the  above  table.  Before  using 

other cleaning methods, the user must confirm with the 

manufacturer  /  supplier  that  the  method  will  not 

damage  the  instrument.  When  cleaning  the  product, 

please wear suitable protective gloves. 

Содержание SCI100-S

Страница 1: ...a at t to o or r r Operation Instructions LED Digital Rotary Evaporator Please carefully read the Instructions and use the product safely under the direction of the Instructions If any information of...

Страница 2: ...he use of the Manual the company will not be liable for this at all The manual is a pure technical document free of any implication or oblique hint of any third party Moreover we won t be liable for a...

Страница 3: ...cation 5 3 Inspection 6 3 1 Open package inspection 6 3 2 Package list 6 4 Control and display 6 4 1 Control 6 4 2 Display 9 5 Installation and testing 11 5 1 Installation 11 5 2 Testing 14 6 Fault di...

Страница 4: ...nty card Description of problem phenomenon The method and operating steps you have undertaken to resolve the problem Your contact details including telephone number fax number and email address Warran...

Страница 5: ...obiological medium might cause damage to the operator Place the instrument on a stable clean antiskid dry and fireproof table surface Make sure there is sufficient space above the instrument and the g...

Страница 6: ...rantee the operational safety Please refer to the operating instructions for heating kettle Please refer to the operating instructions for optional accessories Please use this instrument in a closed f...

Страница 7: ...flask and glass assembly might become hot in the operation process and remain hot for a while after stop of the operation Before further operation please let each part cool down Note be careful to av...

Страница 8: ...es are protected from extrusion and collision Only professional repairman may open the instrument 2 Scope of application This instrument is designed for such application environments as schools labs a...

Страница 9: ...e contact supplier or manufacturer Caution 1 If you find any visible damage on the product please don t connect the product to power supply 3 2 Package list Name Qty Main unit 1 Heating bath 1 Power c...

Страница 10: ...Rotary Evaporator 7 Fig 2 Fig 3 Main unit three in one switch Heating bath power interface Heating bath power switch...

Страница 11: ...erse rotation interval time Turning the adjustment knob can set the intermittent left and right rotation time within the range of 1 999 sec When set to 0 sec the instrument rotates forward only Adjust...

Страница 12: ...es the real time speed Fig 6 This interface indicates the real time speed Turn the adjustment knob to set the speed and press the adjustment knob to start the instrument The rotation speed can be chan...

Страница 13: ...get temperature of the heating bath After setting by pressing the knob the heating indicator is on and the heating process begins In the heating process the target temperature can be changed by turnin...

Страница 14: ...Fig 11 Place the main unit on a flat tabletop and squeeze the handle switch a to lift it to the highest position Fig 12 Tighten the screw b by hand adjust the angle of the motor head and tighten the...

Страница 15: ...mount the evaporating tube in place and rotate the locking device d clockwise for locking Mount the seal ring Fig 13 and Fig 14 Fig 14 Fig 15 Mount the condenser f Fig 15 and Fig 16 Slide the locking...

Страница 16: ...ry Evaporator 13 Fig 16 Fig 17 Mount the fixing device for the vertical condensation glass assembly Fig 17 Fig 18 Piping connection Fig 18 f Condensate inlet Vacuum tube interface g h g Condensate out...

Страница 17: ...90 under the water bath mode Press the temperature adjustment knob to start heating Squeeze the handle switch to place the evaporating flask into the heating bath Press the adjustment knob to start ro...

Страница 18: ...supply until all surfaces are dry If liquid or moist solid enters the product please disconnect the power supply quickly and leave off and contact the manufacturer supplier for more advice Keep the p...

Страница 19: ...ower W 45 1010 Motor DC brushless Speed range rpm 20 200 Display LED Temperature range C Room temperature 180 Temperature control accuracy C 1 oil 3 Lifting method Manual Lifting travel mm 110 Forward...

Страница 20: ...Rotary Evaporator 17 External dimension W x D x H mm 440 320 450 300 300 240 Weight kg 7 3 Allowable ambient temperature C 5 40 Allowable RH 80 DIN EN60529 protection class IP20 USB Yes...

Страница 21: ...SCILOGEX LLC 1275 Cromwell Ave Suite C6 Rocky Hill CT 06067 USA Tel 1 860 436 9221 Fax 1 860 436 9745 info scilogex com www scilogex com...

Отзывы: