6
Tube Roller
5. Operation
•
Place the Rotator in safe and stable surface, ensure the
required operating voltage and power supply voltage
matched, and the socket must be earthed reliably.
•
Connect the power cable.
•
Switch ON instrument.
•
The instrument begins self-checking.
•
When initialization is over, LCD displays the last
running values of speed.
•
Rotate the speed/timer control knob to set target speed.
a.
Press the speed/timer control knob to start
rotating function. Press the speed/timer control knob
again to stop rotating function.
b.
Press the mode switch button, LCD displays time,
and then rotating the speed/timer control knob to set
target time. Press the speed/timer control knob to start
rotating function at the set time. During operation,
motion can be stopped at any time by pressing the
speed/timer control knob. If the knob is pressed again,
motion will start again and the timer will restart
countdown. When the timer reaches zero, the unit will
be automatically halted and an alert will sound.
If these operations above are normal, the instrument is
ready to operate. If not, the instrument may be damaged
during transportation, please contact technical support of
manufacturer/supplier.
Note:
The set values can be altered at any time.
6. Faults
•
Instrument cannot be powered ON when start rotating
function.
n
Check whether the power cable is connected.
•
Rotating function suddenly stop
n
Power/overload protection light changed to red,
LCD displays “Er 03”, indicate the current failure
is “overload protection”.
n
Pressing the speed/timer control knob first to
close rotating function, and then restart rotating
function after lowered load. If rotating function
cannot start, then repeat the process and gradually
reduce load.