background image

 

 

9

BlueSpin

 (Hotplate) Magnetic Stirrer

 

Hotplate center and outer edge. 

 

The sample in the vessel and hotplate surface. 

In order to ensure the accuracy of the temperature inside the 
container, please use the external temperature sensor 
PT1000. 

6.1 Working with external temperature 

sensor 

 

Figure 6 

The external temperature sensor PT1000 is the 
manufacturer’s standard accessory. When PT1000 sensor is 
connected and rotate the temperature control knob, LED 
displays the temperature setting value and shifts to real 
value in 5 seconds. Safe circuit controls hotplate 
temperature under 320°C. Comparing with the temperature 
control of the hotplate, the external temperature sensor can 
control the medium’s temperature more precise. The 

heating function will be stopped automatically under 
abnormal conditions. Please operate follow the instructions 
below: 

 

Switch OFF the instrument. 

 

Ensure the external temperature sensor is inserted in 
the media heated. 

 

Switch ON the instrument and run heating function. 

If the heating function does not work, please contact 
manufacturer/supplier for technical support. 

6.2 Residual heat warning

HOT

 

In order to prevent the risk of burns from a hotplate, digital 
hotplate has a residual heat warning function. When the 
heating function is switched off and the heating plate 
temperature is still above 50°C, “Hot” will flash to warn 
that there is a hazard of burns from the hotplate. When the 
unit is powered off, the LED screen displays the 
temperature of hotplate and Hot in turn. When the hotplate 
temperature drops to below 50°C, the unit will 
automatically switch off. If users want to turn off the LED 
immediately, just pull out the plug directly. When the plug 
is pulled out, the residual heat warning function cannot be 
run. 

Содержание MS-H280

Страница 1: ...H280 Pro BlueSpin LED Digital Hotplate Magnetic Stirrer MS PA BlueSpin LED Digital Magnetic Stirrer MS PB BlueSpin Classic Magnetic Stirrer Please read the User Manual carefully before use and follow all operating and safety instructions www scilogex com 12300175 ...

Страница 2: ...tems 4 4 Control 4 4 1 Control elements 4 4 2 Display 7 5 Trial Run 8 6 Function Heating Digital hotplate model 8 6 1 Working with external temperature sensor 9 6 2 Residual heat warning HOT 9 7 Stirring 10 8 Faults 10 9 Maintenance and Cleaning 11 10 Associated Standards and Regulations 11 11 Specifications 12 12 Ordering Information 14 ...

Страница 3: ...ng information Serial number on the rear panel Certification Description of problem i e hardware or software Methods and procedures adopted to resolve the problems Your contact information Warranty This instrument is warranted to be free from defects in materials and workmanship under normal use and service for a period of 24 months from the date of invoice The warranty is extended only to the ori...

Страница 4: ...res with hazardous substances or under water Gradually increase the speed reduce the speed if Stirring bar breaks away due to high speed The instrument is not running smoothly or container moves on the base plate Temperature must always be set at least 50 lower than the fire point of the media used Be aware of hazards due to Flammable materials or media with a low boiling temperature Overfilling o...

Страница 5: ...s not touch the hotplate Do not cover the device The instrument may only be opened by experts Keep away from high magnetic field 2 Proper Use The instrument is designed for mixing and or heating liquids in schools laboratories or factories Observe the minimum distances between the devices between the device and the wall and above the assembly min 100 mm Figure 1 This device is not suitable for usi...

Страница 6: ...er for technical support Note If there is any apparent damage to the system please do not plug it into the power line 3 2 Listing of Items The package includes the following items Items Qty Main unit 1 Power cable 1 Stirrer bar 1 User Manual 1 Table 1 4 Control 4 1 Control elements Speed Control Knob LED Display Temperature Control Knob Hotplate model Probe Hotplate model Figure 2 Digital hotplate...

Страница 7: ...5 BlueSpin Hotplate Magnetic Stirrer Speed Control Knob LED Display Figure 3 Digital model Speed Control Knob Figure 4 Classic model Figure 5 ...

Страница 8: ...l working state and all settings Probe When the external temperature sensor PT1000 is plugged in probe icon is lit Power Switch Switch ON or OFF the instrument Digital model Temperature Control Knob Stir Set the rotary speed The function Stirring is switched ON or OFF via push ON OFF knob LED Display LED displays the real working state and all settings Power Switch Switch ON or OFF the instrument ...

Страница 9: ...tplate temperature is still above 50 C LED displays Hot otherwise LED displays OFF Speed display area When stirring function was switched ON LED displays the speed setting value and flashes The setting value does not flash until real speed reaches the set point Table 3 Figure 7 Digital model Characters Descriptions Speed display area When stirring function was switched ON LED displays the speed se...

Страница 10: ...rking 6 Function Heating Digital hotplate model The device is controlled by digital temperature control technology which has two separate safe circuits The hotplate is kept at a constant temperature by a digital control circuit The hotplate temperature can also be monitored from a separate adjustable safe circuit The two temperature sensors internal for temperature control are built into the hotpl...

Страница 11: ...ly under abnormal conditions Please operate follow the instructions below Switch OFF the instrument Ensure the external temperature sensor is inserted in the media heated Switch ON the instrument and run heating function If the heating function does not work please contact manufacturer supplier for technical support 6 2 Residual heat warning HOT In order to prevent the risk of burns from a hotplat...

Страница 12: ... power ON Check whether the power line is unplugged Check whether the fuse is broken or loose Fault in power on self test Switch OFF the unit then switch ON and reset the instruments to factory default setting Stir speed cannot reach set point Excessive medium viscosity may cause abnormal speed reduction of the motor Unit cannot be powered off when switched off Check if the residual heat warning f...

Страница 13: ...nufacturer that this method will not harm the instrument Wear the proper protective gloves during cleaning of the instrument Note Electronic device can not clean with cleanser If you require maintenance service must be cleaned the instrument in advance to avoid pollution of hazardous substances and to send back into original packing If the instrument will not use for a long time please switch off ...

Страница 14: ...requency energy and if not installed and used in accordance with the instruction manual may cause harmful interference to radio communications Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense 11 Specifications Items Specifications Digital model Classic model Voltage VAC 1...

Страница 15: ...LED Temperature display accuracy C 1 The safe temperature of hotplate 320 Temperature sensor in medium PT1000 Control accuracy of heating temperature with temperature sensor 0 5 Residual heat warning 50 Dimensions mm 220 160 95 Items Specifications Digital model Classic model 220 160 75 Weight kg 1 4 0 7 0 7 Permitted ambient temperature C 5 40 Permitted relative humidity 80 Protection class acc t...

Страница 16: ...9999 MS H280 Pro BlueSpin Double LED Digital Hotplate Magnetic Stirrer stainless steel with corrosion resistant coated hotplate USA plug 100 120V 50Hz 60Hz 861131129999 MS H280 Pro BlueSpin Double LED Digital Hotplate Magnetic Stirrer stainless steel with corrosion resistant coated hotplate Cn plug 200 240V 50Hz 60Hz 861131229999 MS H280 Pro BlueSpin Double LED Digital Hotplate Magnetic Stirrer st...

Страница 17: ...gth 230mm 18900136 PT1000 B Temperature sensor with glass coated 230mm 18900148 Support clamp of PT1000 18900080 Reaction block for 50ml round bottom flask one flask capacity 18900081 Reaction block for 100ml round bottom flask one flask capacity 18900082 Reaction block for 250ml round bottom flask one flask capacity 18900083 Reaction block for 500ml round bottom flask one flask capacity 18900006 ...

Отзывы: