Description
Item
Connect to a protective earth terminal or other grounding point of the module. (Preventive
Measure One)
CAUTION: Potential System Damage. Do not connect the grounding
wire to the gas tubing, the water service tubing, or the telephone line.
4
Static electricity generated by the liquid will be conducted through the container to the
earth.
5
Preventive Measure One
Use a metal container for the waste liquid, and connect the container to protective earth to make sure that the
electrical charges of the container and liquid pass to the earth.
Required Materials (not provided)
• Grounding wire with clip
• 18 L metal container
• 4 L metal container
Guidelines
• Connect the metal waste container to protective earth properly. If the grounding wire is not properly connected
to protective earth, then static electricity can build up in the container.
CAUTION: Potential System Damage. Do not connect the grounding wire to the gas
tubing, the water service tubing, or the telephone line.
Note:
Some metal containers have surfaces that are laminated or oxidized, and therefore do not conduct
electricity. After connecting the metal container to protective earth, use a tester to verify that electricity is
conducted to the earth.
• If the liquid to be drained into the waste container is virtually nonconductive (10
–10
S/m or less), then add
properly conductive liquid to the tank. This conductive liquid can be added beforehand.
Preventive Measure Two
Cover the spaces between the tubing and the sides of the inlet and outlet openings of the waste container with
caps or other protective covering. This prevents any sparks generated outside the container from getting inside.
ExionLC
™
100 System
Hardware User Guide
15 / 219
RUO-IDV-05-1839-C
Operational Precautions and Limitations