Before the Installation
Before the CESI 8000 Plus system can be installed, the mass spectrometer must be operational. To demonstrate
this, a performance test should be performed and a printout of the results attached to this document.
For some mass spectrometers, additional equipment is required, refer to the appropriate section below.
During the Installation
During the installation, a person who is familiar with the operation of the mass spectrometer should be available
to operate the system.
SCIEX Mass Spectrometers
For the OptiMS Adapter for SCIEX NanoSpray III source (PN B07363), all systems require the NanoSpray
®
III ion
source retrofit kit (PN 5030793), purchased from SCIEX.
Make sure that the required orifice plate for the NanoSpray
®
ion source is present (refer to the following table).
Orifice Plate
Part Number
NanoSpray
®
III Ion
Source Part Number
Model
1033731
5028129
SCIEX Triple Quad
TM
5500 system
QTRAP
®
5500 system
TripleTOF
®
4600, 5600, and 5600+ systems
1033731
5029154
TripleTOF
®
6600 system
5018950
5028138
SCIEX Triple Quad
TM
6500 and 6500+ systems
QTRAP
®
6500 and 6500+ systems
Thermo Scientific Mass Spectrometers
For the OptiMS Adapter for Thermo Nanospray II MS Sources (PN B07366), make sure the additional required
equipment is present (refer to the following table).
Additional Required Equipment
Ion Source
Model
Ion MAX adapter
Nanospray II ion source
LCQ Deca XP MAX, LTQ series and the TSQ
series systems
Adapter ring
Nanospray II ion source
LCQ Deca XP and LCQ Advantage XP systems
For the OptiMS Thermo MS Adapter for Nanospray Flex and Nanospray Flex NG Sources (PN B83386), no additional
equipment is required.
CESI 8000 Plus High Performance
Separation-ESI Module
Site Planning Guide
19 / 21
RUO-IDV-05-3345-C | B11953AF