• Relative humidity from 20% to 60%, non-condensing
BioSafety Requirements
The site must not be designated BioSafety Level 3 (BSL-3) or BioSafety Level 4 (BSL-4). SCIEX does not install,
service, or repair SCIEX systems in areas designated BSL-3 or BSL-4.
Computer, Software, and Network Requirements
A computer (also referred to as a "controller") and monitor are supplied with the CESI 8000 Plus High Performance
Separation-ESI Module system.
Refer to the
Site Planning Guide
for the mass spectrometer for detailed requirements for the mass spectrometer
acquisition computer.
Controller Requirements
The controller and monitor are provided with the CE system.
CAUTION: Potential System Damage. Other than security software, do not install additional
software on the controller. Changes to the configured software could void the system
warranty and cause the system to stop operating. All operators require read/write access
to the C:\32Karat folder.
Printer Requirements
The system can be connected to a network or dedicated printer.
LAN Connection (Optional)
The controller can be connected to the network.
Customer-Supplied Solutions and Equipment
WARNING! Toxic Chemical Hazard. Refer to the chemical product Safety Data Sheets
and follow all of the recommended safety procedures when handling, storing, and
disposing of chemicals. For health and safety precautions, refer to the User Guide.
CESI 8000 Plus High Performance
Separation-ESI Module
Site Planning Guide
17 / 21
RUO-IDV-05-3345-C | B11953AF
Site Requirements