Table A-1 CESI 8000 Plus High Performance Separation-ESI Module Electrical Specifications
Value
Specification
Instrument: 100 VAC to 240 VAC, auto-ranging
Cart: 100 VAC to 240 VAC, auto-ranging
Controller: 100 VAC to 240 VAC
Monitor: 100 VAC to 240 VAC
Nominal input voltage
50 Hz or 60 Hz
Frequency
Instrument: 8 A
Cart: 4.5 A
Controller: 2.8 A
Monitor: 1.5 A
Maximum input current
Ventilation and Waste Collection Requirements
Ventilation Requirements
Refer to the "Laboratories" chapter in
ASHRAE, 2011 Handbook - HVAC Applications
for recommended
laboratory ventilation guidance.
Waste Collection Requirements
WARNING! Biohazard, Toxic Chemical Hazard. Follow local directives when
disposing of chemicals, vials and caps, and the remains of the prepared
samples, if applicable. They might contain regulated compounds and
biohazardous agents.
Environmental Requirements
• Altitude not exceeding 2000 m (6562 feet) above sea level
• An ambient temperature of 15 °C to 30 °C (59 °F to 86 °F)
Site Planning Guide
CESI 8000 Plus High Performance
Separation-ESI Module
RUO-IDV-05-3345-C | B11953AF
16 / 21