3. Click
(Service) and then click the
Maintenance
tab.
4. Click
Instrument ID
.
5. Type the serial number in the
Instrument ID
field and then click
Apply
.
Install the Cartridge in the System
CAUTION: Potential System Damage. Handle the cartridge with care. The tips of the
capillaries are made of glass and are fragile. If the tips are broken, then replace the cartridge.
1. If the cartridge has been stored in the refrigerator, it must come to room temperature before use. Follow these
instructions:
a. Remove the cartridge from the storage box and then put it in the cartridge stand.
b. Add enough DDI water to the cartridge stand trough to keep the capillary tips submerged so that they do
not dry out.
c. Allow the cartridge to stand at room temperature for at least 20 minutes before use.
2. Remove the cartridge from the cartridge stand.
3. Remove the seal from the cartridge purge port and then, using a laboratory tissue, wipe off any gel residue.
Refer to
Store the seal in a safe location for later use when storing the cartridge.
4. Open the cartridge door and then put the cartridge in the C100
HT
Biologics Analyzer system so that the label
faces the front of the system.
5. Put the data key in the socket in either orientation and then close the cartridge door.
If the
Latch cartridge automatically
option is selected in the C100
HT
software Settings, then the cartridge
automatically engages and the optical and pressure systems are linked to the cartridge.
C100
HT
Biologics Analyzer System
Installation Guide
15 / 31
RUO-IDV-09-3829-A
Installation Instructions