Ion Source Optimization
4
WARNING! Ionizing Radiation Hazard, Biohazard, or Toxic Chemical
Hazard. Do not use the ion source without knowledge of and training
in the proper use, containment, and evacuation of toxic or injurious
materials used with the ion source.
WARNING! Fire Hazard. Do not direct more than 3 mL/min of flammable solvent in
the ion source. Exceeding the maximum flow rate can cause solvent to accumulate
in the ion source. Do not use the ion source if the source exhaust system is not
enabled and functioning when the ion source and the probe are properly installed.
WARNING! Puncture Hazard, Ionizing Radiation Hazard, Biohazard, or
Toxic Chemical Hazard. Discontinue use of the ion source if the ion
source window is cracked or broken, and then contact a SCIEX Field
Service Employee (FSE). Any toxic or injurious materials introduced in
the equipment will be present in the source exhaust output. Exhaust from
equipment should be vented from the room. Dispose of sharps following
established laboratory safety procedures.
Optimize the ion source whenever the analyte, flow rate, or mobile phase composition changes.
When optimizing ion source-dependent parameters, introduce the sample at a flow rate that will
be used during sample analysis, using either flow injection analysis (FIA) or tee infusion as the
method of sample introduction. Optimize the position of the ion source before optimizing the ion
source-dependent parameters.
Several parameters affect the performance of the source. Optimize the performance while
injecting a known compound and monitoring the signal of the known ion. Adjust the micrometer
and gas and voltage parameters to maximize the signal-to-noise ratio and signal stability.
Refer to the section:
TurboIonSpray Probe Optimization
.
Sample Introduction
Method
The liquid sample stream is delivered to the ion source by an LC pump or by a syringe pump. If
it is delivered by an LC pump, then the sample can be injected directly in the mobile phase using
Ion Source Optimization
System User Guide
RUO-IDV-05-6475-C