![SCIEX 3200 QTRAP Скачать руководство пользователя страница 14](http://html1.mh-extra.com/html/sciex/3200-qtrap/3200-qtrap_system-users-manual_1217527014.webp)
Physical Precautions
WARNING! Lifting Hazard. Use a mechanical lifting device to lift and move the
mass spectrometer. If the mass spectrometer must be moved manually, then at
least six people are required to move it safely. Follow established safe lifting
procedures. We recommend the use of a professional moving service. For the
weights of system components, refer to the document: Site Planning Guide.
Environmental Precautions
Use qualified personnel for the installation of electrical mains, heating, ventilation, and plumbing
supplies and fixtures. Make sure that all of the installations comply with local bylaws and
biohazard regulations. For information about the required environmental conditions for the
system, refer to the document:
Site Planning Guide
.
Allow access space around the equipment when setting up the system.
DANGER!
Explosion Hazard. Do not operate the system in an environment containing
explosive gases. The system is not designed for operation in an explosive environment.
WARNING! Biohazard. For biohazardous material use, always comply with local
regulations for hazard assessment, control, and handling. This system or any part
is not intended to act as a biological containment.
WARNING! Environmental Hazard. Follow established procedures for disposal
of biohazardous, toxic, radioactive, and electronic waste. The customer
is responsible for disposal of hazardous substances, including chemicals,
waste oils, and electrical components, in accordance with local laws and
regulations.
CAUTION: Potential Mass Shift. Maintain a stable ambient temperature. If the temperature
changes by more than 2 °C per hour, then the resolution and mass calibration might be
affected.
Electromagnetic Environment
Electromagnetic Compatibility
Basic Electromagnetic Environment:
Environment existing at locations characterized by being
supplied directly at low voltage from the public mains network.
Operational Precautions and Limitations
SCIEX 3200 Systems
14/114
System User Guide
RUO-IDV-05-6475-C