Error or Indicator
Cause
Explanation / Solution
ZEROING is displayed for an extended
period.
The initializing process was
interrupted
Don’t move the force gauge and press
0
ENTER key.
AD range exceeded (+/-)
The initializing process interrupted.
Possible overload or load-cell
damage
Put the gauge in horizontal position and turn it
off and on using the
ON/OFF
key. Contact
technical support.
The values indicated by the force
gauge differs significantly from the
reference.
Calibration may be required.
Contact a servicing facility to calibrate the
gauge.
Chapter 19: Common Errors and Troubleshooting
Chapter 24: MaintenanceChapter 16: Common Errors and
Troubleshooting