67
Instruction Manual
– FB and FC Series Force Gauges
The threshold output mode enables the Output port by sending voltage level signals which can be used for signaling or controlling
external peripherals that can connect with the force gauge. The output mode can be used in combination with Programmable
Thresholds Monitoring. Input mode allows external button or a pedal to be used as trigger for starting force measurements and
taking samples.
OUTPUT Ampacity: I
max
=25mA / U
nom
=24V (open collector type, emitters connected
–
GND)
INPUT Voltage Range IN (+)/IN (-): U
in
=12-18V / I
in max
=50mA
Outputs MAX and ZERO: Equivalent (common GND)
Recommended Power: 24V
DC
MIN, MAX, ZERO Output Current: I
out max
= 25mA
Supply voltage: U
in
=3 ÷ 6V
Input current: I
in
=10÷20mA (I
in max
=50mA)
Chapter 18: Input / Output Port Configuration
(FC Models Only)
Chapter 14: Input / Output Port Configuration
(FC Models Only)
Содержание TORBAL FB Precision Series
Страница 1: ...Instruction Manual TORBAL FORCE GAUGES FB Precision FC Precision PRO Series...
Страница 75: ...73 Instruction Manual FB and FC Series Force Gauges...
Страница 76: ...74 Instruction Manual FB and FC Series Force Gauges...
Страница 77: ...75 Instruction Manual FB and FC Series Force Gauges...
Страница 78: ...76 Instruction Manual FB and FC Series Force Gauges...
Страница 79: ...77 Instruction Manual FB and FC Series Force Gauges...
Страница 80: ...78 Instruction Manual FB and FC Series Force Gauges...
Страница 81: ...79 Instruction Manual FB and FC Series Force Gauges...
Страница 82: ...80 Instruction Manual FB and FC Series Force Gauges...
Страница 83: ...81 Instruction Manual FB and FC Series Force Gauges...
Страница 85: ...Scientific Industries Inc 80 Orville Dr Bohemia NY 11716 Tel 973 473 6900 Fax 973 777 8302 www torbalscales com...