3 Operation
Page 12
STAT
MATIC
smart 1,2,3
EN
3
Operation
Manual pre-cleaning of instruments
•
Clean the instrument primarily in the head area with a medium-hard toothbrush for 20 sec. under running water (non
strong water jet) until the instrument looks fully clean. (drinking water, temperature 30 °C
5 °C).
•
Leave the instrument for 5 minutes.
"Cl Maintenance" program - for regular and contra-angle handpieces and turbines
Turn on the mains switch.
LED
(Stand-by) lights up green. "Cl Maintenance" will appear on the
display.
Open the front door.
Place regular and contra-angle handpieces and turbines on the adapters provided
for this purpose (instruments must snap into place). A frame will appear on the
display around the station number per occupied slot.
Close the front cover.
Push the
(Program Start) button on the control membrane keyboard.
Program runs automatically. During the "Cl Maintenance" program,
a clock will count backwards until 00:00 on the display.
All attached regular and contra-angle handpieces and turbines are cleaned and
maintained for inside in turn. After a successful process, the program slots will be
displayed inversely. The display shows: "Cl Maintenance successful".
Open the front door.
Remove regular and contra-angle handpieces and turbines by pressing the
locking button (A). Pull off turbines downwards (press support lever down).
The frame around the slot number will disappear from the display.
Wipe off the instruments using a damp cloth soaked with disinfectant. For Germany we recommend using disinfectants
listed by the VAH (such as SciCan OPT
IM
BLUE).
•
Immediately following this treatment, the instrument must be treated further in accordance with the manufacturer's
instructions on reprocessing (such as disinfection/sterilization in the STAT
IM
).
•
Always perform a function test before using the instruments on patients. Some liquid may come out.
"Collet care" program
Turn on the mains switch.
LED
(Stand-by) lights up green. "Cl Maintenance" will appear on the
display.
Open the front door.
Push instrument with chuck opening into the collet care nozzle.
Activate
(Collet care) key on the control membrane keyboard.
Program runs automatically. A clock on the display runs backwards until 00:00
The following is then shown on the display: " Collet care measure successful".
Remove instrument from the nozzle.
•
Wipe off the instruments using a damp cloth soaked with disinfectant. For Germany we recommend using disinfectants
listed by the VAH (such as SciCan OPT
IM
BLUE).
•
Immediately following this treatment, the instrument must be treated further in accordance with the manufacturer's
instructions on reprocessing (such as disinfection/sterilization in the STAT
IM
).
•
Always perform a function test before using the instruments on patients. Some liquid may come out.