2 Commissioning
Page 11
STAT
MATIC
smart 1,2,3
EN
Setting the date/time
Turn on the device
Press the
button. "Export log data" will appear on the display.
Press the
button 3x until "Date/time" is highlighted in black on the display.
The "Date/time" menu is selected using the
button.
Set the desired value using the
and
buttons
Using the
button, the cursor is moved to the next possible setting. The values are set in the following order: Year,
month, day, hour, minute.
After final actuation using the
, the program jumps back again.
Press the
button to return to the "Cl Maintenance" main menu.
Assemble and check the adapter
Assemble the adapter, if required
Check the MULTIflex receptacle (and/or other turbine adapter) for firm seating.
The clamping nut must be tight.
Retighten clamping nut, if required.
Please note that only white STAT
MATIC
smart / STAT
MATIC
PLUS adapters can be
used. Adapters (black) of the STAT
MATIC
maintenance unit are not suitable.
Check the output quantity on all cleaning slots using a testing adapter intended
for the coupling.
Functional testing of flow rate using the testing adapter
Attach the matching testing adapter to the corresponding instrument couplings
(such as INTRAmatic, DIN ISO EN coupling).
Close the front cover.
Press the START button.
After completion of the program, check the filling quantity in each of the testing
adapters.
The fluid must be above the mark ("Min.") !
If the filling quantity is below the mark, please consult the SciCan customer
service.
Please note that only STAT
MATIC
smart testing adapters can be used. STAT
MATIC
PLUS testing devices are not suitable.
After this process has completed, the STATMATIC smart system is operational.